Agilent Technologies 66309B, 66311B, 66111A manual Procedure Display Explanation

Page 46

4 - Turn-On Checkout

 

 

 

 

Procedure

Display

Explanation

7.

Plug the output connector

 

 

Restores the output sense connections. The Prot

 

back into the unit.

 

 

annunciator is still on.

8.

Press Shift, Prot Clear

NO FAULT

Clears the protection condition. Prot is off; CV is on.

9.

Press Voltage

VOLT 0.000

Display shows the output voltage setting of the unit.

10.

Press Enter Number,

VOLT <15>

Programs the main output to 15 volts. After the value

 

<15>, Enter

15.003V

0.0001A

is entered, the display returns to Meter mode.

 

 

 

 

Because the output is enabled, the meter will indicate

 

 

 

 

the actual output voltage.

11.

Press Shift, OV

PROT:LEV 22.00

Display shows the overvoltage protection trip voltage

 

 

 

 

of the unit.

12.

Press Enter Number,

PROT:LEV <8>

Programs the OVP to 8 volts, which is less than the

 

<8>, Enter

 

 

previously set output voltage.

 

 

0.000V

0.0001A

Because the OVP voltage entered was less than the

 

 

 

 

output voltage, the OVP circuit tripped. The output

 

 

 

 

dropped to zero, CV turned off, and Prot turned on.

13.

Press Shift, OV,

PROT:LEV <22>

Programs the OVP to a value greater than the output

 

Enter Number,

 

 

voltage setting of the unit. This prevents the OV

 

<22>, Enter

 

 

circuit from tripping again when the protection

 

 

 

 

condition is cleared.

14.

Press Shift, Prot Clear

15.003V

0.0001A

Clears the protection condition, thus restoring the

 

 

 

 

output of the unit. Prot turns off and CV turns on.

15.

Press Output On/Off

0.000V

0.0000A

Turn the output off.

16.

Connect a jumper wire

 

 

Shorts the output of the unit.

 

across the + and - output

 

 

 

 

terminals.

 

 

 

17.

Press Output On/Off.

0.004V

3.0712A

The CC annunciator is on, indicating that the unit is

 

 

 

 

in constant current mode. The unit is sourcing output

 

 

 

 

current at the maximum rating, which is the default

 

 

 

 

output current limit setting.

18.

Press Current,

CURR <1>

Programs the output current to 1 ampere.

 

Enter Number,

 

 

 

 

<1>, Enter.

 

 

 

19.

Press Shift, OCP

0.001V

0.0003A

You enabled the overcurrent protection circuit. The

 

 

 

 

circuit then tripped because the unit was operating in

 

 

 

 

constant current mode. The CC annunciator turns off,

 

 

 

 

and the OCP and Prot annunciators turn on.

20.

Press Shift, OCP

0.001V

0.0003A

You have disabled the overcurrent protection circuit.

 

 

 

 

The OCP annunciator turns off.

21.

Press Shift, Prot Clear

0.004V

0.998A

Restores the output. The Prot annunciator turns off.

 

 

 

 

The CC annunciator turns on.

22.Turn the unit off and remove the shorting wire from the output terminals.

The next time the unit turns on it will be restored to the *RST or factory default state.

46

Image 46
Contents USER’S Guide Warranty Information CertificationSafety Summary GeneralSafety Symbols Declaration EMCDeclaration of Conformity Acoustic Noise Information Printing HistoryTable of Contents Types of Scpi Commands Checkout Procedure Case of TroubleVXIplug&play Power Products Instrument Drivers DVM ConnectionsIntroduction Programming the Output Types of Scpi MessagesScpi Data Formats Triggering Output ChangesDisplay Commands 103 Inhibit/Fault IndicatorCalibration Commands 100 Measurement Commands 104System Commands 125 Output Commands 114Status Commands 121 Trigger Commands 126Specifications 143 Supplemental Characteristics 144 Common CommandsAdditional Commands Performing the Calibration Procedure 154Basic Error Number List 161Introduction 165 Introduction 177Quick Reference Front Panel At a GlanceUse the front panel Address key to configure the interface Instrument ConfigurationRear Panel At a Glance Front Panel Number Entry Use the Function keys and Entry keys to enter a new valueFront Panel Annunciators Immediate Action KeysFront Panel Menus At a Glance Scpi Programming Commands At a Glance ABORt CALibrateProgramming the unit using Scpi COMPatibility commands Using the front panelUsing the programming interface Installing the VXIplug&play instrument driverOption Safety ConsiderationsOptions and Accessories DescriptionAgilent 66309B Description and Model DifferencesAgilent Agilent 66309DFront Panel Controls Common CapabilitiesRemote Programming Output 1 Characteristic Dc Source Output 1 CharacteristicOutput 2 Characteristic Output 2 CharacteristicOption 521 Description Agilent 66309B/D only Option 521 Factory SettingsOption 521 Relay Modes Page Additional Agilent 66311/66309 Operating Settings Checks Installation and Operation ChecklistCheck the Operating Settings and Conditions Check the Phone ConnectionsLocation CleaningInspection Damage Packaging MaterialBench Operation Connect the Power CordInput Connections Rack MountingOutput Turn the unit off before connecting any wiresOutput Connections Current RatingsRemote Sense Connections Remote Sense ConnectionsRemote Sense Connections with External Relays Load Regulation and Voltage Drop in the Remote Sense Leads Maintaining Stability while Remote SensingOpen Sense Lead Protection Local Sensing Message DescriptionOutput Compensation DVM Connections OVP ConsiderationsMeasuring Circuits that are Not Powered by the Main Output Measuring Circuits Not Powered by the Main OutputMeasuring Circuits Floating with Respect to the Main Output External Protection Connections 10. FLT/INH ExamplesComputer Connections FLT/INH DIGital I/O ConnectorDigital I/O Connections Gpib InterfaceRS-232 Interface Pin Input/Output DescriptionPress Output On/Off Checkout ProcedureProcedure Display Explanation Press ProtectProcedure Display Explanation Press Shift, Channel Selftest Error MessagesCase of Trouble Enter Number 12, Enter Press Output On/OffRuntime Error Messages Runtime Error MessagesPower-On Selftest Errors Line FuseIntroduction Front Panel DescriptionFront Panel Operation System Keys Display Command FunctionOFF Immediate Action Keys Function KeysScrolling Keysq Metering Keys Display MeasurementOutput Control Keys Over CurrentEntry Keys Entry KeysSelecting an output on Agilent 66309B/D units Using the Front Panel DisplayExamples of Front Panel Programming Selecting the DVM on Agilent 66311D/66309D unitsSet the output compensation Set the output voltageSet the output current limit Typecap HighSet the output 2 current limit Enable the outputSet the output 2 voltage Keypad, press Enter Number, 7, EnterQuery and Clear Errors as follows Querying and Clearing Output Protection and ErrorsDisable Overvoltage Protection as follows OvercurrentMaking Basic Front Panel Measurements Use the Meter menu for making front panel measurementsDefault Front Panel Measurement Parameters Making Enhanced Front Panel MeasurementsCurrrang Auto Making DVM Measurements Agilent 66311D/66309D onlyUse the Meter menu for making DVM measurements Currdet AcdcTo configure the Digio mode of the port, proceed as follows Setting the Gpib Address and Programming LanguageTo configure the Ridfi mode of the port, proceed as follows Set the Gpib address as followsStoring and Recalling Instrument States Gpib References External ReferencesScpi References Supported Applications VXIplug&play Power Products Instrument DriversDownloading and Installing the Driver System RequirementsRS-232 Capabilities of the DC Source Accessing Online HelpGpib Capabilities of the DC Source Gpib AddressRS-232 Flow Control Introduction to ScpiBaud Rate Conventions Used in This Guide Types of Scpi CommandsBoldface font Including Common Commands Multiple Commands in a MessageMoving Among Subsystems Message Unit Using QueriesTypes of Scpi Messages HeadersScpi Data Formats Response Data Types Scpi Command CompletionSuffixes and Multipliers Class Suffix Unit Unit with MultiplierNon-SCPI Commands Using Device ClearScpi Conformed Commands Scpi Conformance InformationProgramming the Output Power-on InitializationEnabling the Output Output Voltage Output CurrentScpi Triggering Nomenclature Setting the Voltage or Current Transient LevelsTriggering Output Changes Output Trigger ModelGenerating Triggers Enabling the Output Trigger SystemSelecting the Output Trigger Source Single TriggerAverage Measurements Making Basic MeasurementsControlling Measurement Samples Making Enhanced Measurements Window FunctionsCurrent Ranges and Measurement Detector RMS MeasurementsMinimum and Maximum Measurements Pulse MeasurementsHigh/Low Measurements Making DVM Measurements Returning All Measurement Data From the Data BufferMeasurement Trigger Model Triggered MeasurementsSequence Form Alias SEQuence2 ACQuire Selecting the Sensing Function Enabling the Measurement Trigger SystemSelecting the Measurement Trigger Source INTernalGenerating Measurement Triggers Single TriggersTrigacqcouncurr number or Trigacqcounvolt number Programming the Status Registers Pre-trigger and Post-trigger Data AcquisitionPower-On Conditions DC Source Status ModelBit Configurations of Status Registers Operation Status GroupStandard Event Status Group PON Power On BitQuestionable Status Group Status Byte RegisterMSS Bit Determining the Cause of a Service InterruptServicing Operation Status and Questionable Status Events RQS BitMonitoring Both Phases of a Status Transition Inhibit/Fault IndicatorRemote Inhibit RI Bit Weight Discrete Fault Indicator DFIUsing the Inhibit/Fault Port as a Digital I/O PinSubsystem Commands Subsystem Commands Syntax Language Dictionary Language Dictionary Common Commands Programming ParametersCALibrateCURRent2 Calibration CommandsCALibrateCURRent CALibrateCURRentMEASureLOWRangeCALibrateDATE CALibratePASSwordCALibrateDATA CALibrateDVMCALibrateVOLTage CALibrateSAVECALibrateSTATe CALibrateVOLTage2DISPlayCHANnel Display CommandsDISPlay DISPlayMODEFORMat Measurement Commands104 MEASureARRayVOLTage? FETChARRayVOLTage? FORMatBORDerMEASureARRayCURRent? FETChARRayCURRent? Query SyntaxMEASureCURRentACDC? FETChCURRentACDC? MEASureCURRent? FETChCURRent?MEASureCURRent2? NR3MEASureCURRentMAXimum? FETChCURRent MAXimum? MEASureCURRentHIGH? FETChCURRentHIGH?MEASureCURRentLOW? FETChCURRentLOW? 107MEASureDVMACDC? FETChDVMACDC? MEASureCURRentMINimum? FETChCURRentMINimum?MEASureDVM? FETChDVM? MEASureVOLTage? FETChVOLTage?MEASureVOLTageHIGH? FETChVOLTageHIGH? MEASureVOLTage2MEASureVOLTageACDC? FETChVOLTageACDC? 109MEASureVOLTageMINimum? FETChVOLTageMINimum? MEASureVOLTageLOW? FETChVOLTageLOW?MEASureVOLTageMAXimum? FETChVOLTageMAXimum? 110Query Syntax SENSeCURRentDETector? SENSeCURRentDETectorSENSeCURRentRANGe 111SENSeSWEepOFFSetPOINts SENSeFUNCtionSENSePROTectionSTATe SENSeSWEepPOINtsQuery Syntax SENSeSWEepTINTerval? SENSeSWEepTINTervalSENSeWINDow Query Syntax SENSeWINDowTYPE?OUTPut1 Output CommandsINSTrumentCOUPleOUTPutSTATe OUTPut1 2RELayMODEOUTPutPONSTATe OUTPutDFIOUTPutDFISOURce 115OUTPutRIMODE OUTPutPROTectionDELayOUTPutPROTectionCLEar OUTPutTYPESOURceCURRent2 SOURceCURRent117 SOURceCURRentTRIGger Command Syntax Parameters RST Value ExamplesSOURceCURRentPROTectionSTATe SOURceCURRent2TRIGgerSOURceVOLTage SOURceDIGitalDATASOURceDIGitalFUNCtion 119SOURceVOLTagePROTectionSTATe SOURceVOLTage2SOURceVOLTagePROTection 120SOURceVOLTageTRIGger Status CommandsSTATusPRESet SOURceVOLTage2TRIGgerSTATusOPERation? STATusOPERationENABleBit Configuration of Operation Status Registers STATusOPERationCONDition?STATusOPERationNTR STATusOPERationPTR Parameters Preset ValueBit Configuration of Questionable Status Registers STATusQUEStionable?STATusQUEStionableNTR STATusQUEStionablePTR STATusQUEStionableENABleSTATusQUEStionableCONDition? 124SYSTemLANGuage System CommandsSYSTemERRor? SYSTemVERSion?ABORt Trigger CommandsRelated Commands INITiateSEQuence INITiateNAMETRIGgerSOURce TRIGger127 TRIGgerSEQuence2COUNtDVM TRIGgerACQuireCOUNtDVM TRIGgerSEQuence2 TRIGgerACQuireTRIGgerSEQuence2COUNtCURRent TRIGgerACQuireCOUNtCURRent 128Unit TRIGgerSEQuence2COUNtVOLTage TRIGgerACQuireCOUNtVOLTage129 TRIGgerSEQuence2HYSTeresisDVM TRIGgerACQuireHYSTeresisDVM 130TRIGgerSEQuence2LEVelDVM TRIGgerACQuireLEVelDVM TRIGgerSEQuence2LEVelCURRent TRIGgerACQuireLEVelCURRent131 Returned Parameters TRIGgerSEQuence2LEVelVOLTage TRIGgerACQuireLEVelVOLTageTRIGgerSEQuence2SLOPeCURRent TRIGgerACQuireSLOPeCURRent 132TRIGgerSEQuence2SLOPeVOLTage TRIGgerACQuireSLOPeVOLTage TRIGgerSEQuence2SLOPeDVM TRIGgerACQuireSLOPeDVM133 TRIGgerSEQuence1DEFine TRIGgerSEQuence2DEFine TRIGgerSEQuence2SOURce TRIGgerACQuireSOURce134 Bit Configuration of Standard Event Status Enable Register Common CommandsCommand Syntax *CLS Parameters None 135Example Field Information136 Parameters Example Query Syntax *OPT? Returned Parameters AardRCL NRf 137Example *SAV Related Commands *RCL *RST 138 Command Syntax *SAV NRfRST Settings Query Syntax *STB? Power-on ValueBit Configuration of Status Byte Register 139TRG Query Syntax TST? Returned Parameters NR1Additional Commands VOLTageLIMitLOW? CURRentPROTectionTRIPped?VOLTageLIMitHIGH? VOLTagePROTectionTRIPped?Table A-1. Performance Specifications Specifications143 144 Supplemental CharacteristicsTable A-2. Supplemental Characteristics Agilent 66309B/D145 Table A-4. Agilent 66309B/D Option 521 Characteristics 146147 Test SetupEquipment Required Table B-1. Equipment RequiredTurn-On Checkout Performing the Verification Tests148 Output terminals on both outputs 1 Voltage Programming and Measurement AccuracyCurrent Programming and Measurement Accuracy Steps 6-10 apply to Agilent 66309B/D output 2 onlyCurrent Sink Measurement Steps 7-11 apply to Agilent 66309B/D output 2 only150 DVM Measurement Accuracy 151152 153 DVM Voltage Measurement Agilent 66309D onlyFront Panel Calibration Menu Performing the Calibration Procedure154 Enable Calibration Mode Front Panel Calibration ProcedureCheck the Language Setting Voltage Programming and Measurement CalibrationCurrent Programming and High-Range Measurement Calibration Steps 11-16 apply to Agilent 66309B/D output 2 onlyOvervoltage Protection Calibration 156157 Steps 27-32 apply to Agilent 66309B/D output 2 onlyCalcurrmeas AC Saving the Calibration Constants Restore the Language SettingDVM Calibration applies to Agilent 66311D, 66309D only 158Table B-3. Gpib Calibration Error Messages Calibration Error MessagesChanging the Calibration Password Calibration Over the GpibPage Table C-1. Error Numbers Error Number List161 Error Messages 162Error Messages C 163Page National Instruments Gpib Driver Assigning the Gpib Address in Programs165 Example 1. National Instruments Interface Example Error Handling166 167 Example Programs DExample 2. Controller Using Basic 168Example 3. Current Pulse Measurement Using Basic 169170 171 172 DFI Programming Example Example 5. DFI Example Using Basic173 Page Open the Unit Configure the Power TransformerInstall the Correct Line Fuse Close the UnitFigure E-1, Power Transformer AC Input Connections 176Table F-1. COMPatibility Power-on Settings Command Setting177 Table F-2. COMPatibility Commands Compatibility CommandSimilar Scpi Command 178179 Table F-3. COMPatibility Errors ErrorNumberError String Description/Explanation/Examples 180Table F-5. Bit Configuration of Serial Poll Register 181Page Index 183Index 184185 OVERTEMPERATURE, 47 OVERVOLTAGE, 47 OVLD, 47, 60, 61 186Subsystem commands syntax, 96 suffixes 187188 Type CAPEurope Asia Pacific United States Latin AmericaCanada Australia/New Zealand JapanManual Updates
Related manuals
Manual 8 pages 56.17 Kb

66111A, 66309B, 66311B, D specifications

Agilent Technologies D,c,83440b is an advanced electronic measurement solution designed for engineers and scientists who require precise and reliable performance in their testing environments. This modular test system offers a comprehensive suite of features that cater to a wide range of applications, from high-frequency testing to complex signal analysis.

One of the main features of the D,c,83440b is its impressive frequency range, allowing users to conduct tests across a wide spectrum of signals. The system is capable of handling frequencies up to 26.5 GHz, making it ideal for RF and microwave applications. This broad range ensures that users can work with a variety of devices, including communication systems, radar, and satellite technology.

In addition to its frequency capabilities, Agilent Technologies has engineered the D,c,83440b with exceptional dynamic range and low noise figures. This ensures that even the smallest signals can be accurately measured, allowing for greater precision in testing. The full spectrum analysis feature enables users to capture transient events and analyze them in real-time, which is crucial for troubleshooting and performance evaluations.

The D,c,83440b is built on a modular platform, allowing users to customize their systems according to specific testing needs. This modularity not only enhances flexibility but also simplifies maintenance and upgrades. Users can easily swap out different modules without the need for extensive system reconfiguration, which can significantly reduce downtime in testing environments.

Another standout characteristic of the D,c,83440b is its user-friendly interface. With a large, high-resolution display and intuitive controls, engineers can quickly navigate through settings and data, streamlining the testing process. This ease of use is complemented by powerful software solutions that can automate test sequences, aiding in efficiency and accuracy.

The integration of advanced digital signal processing technologies further enhances the capabilities of the D,c,83440b. These technologies enable more sophisticated measurements and improved signal integrity, which is essential for modern communication systems.

In summary, the Agilent Technologies D,c,83440b is a multifaceted electronic measurement solution that boasts a wide frequency range, excellent dynamic range, modular design, and user-friendly interface. This combination of features makes it suitable for various applications, ensuring that engineers and scientists have the tools they need to succeed in their testing and measurement endeavors.