3.8. Initial Calibration Procedure

The next task is to calibrate the analyzer.

To perform the following calibration you must have sources for zero air and span gas available for input into the sample port on the back of the analyzer. See Section 3.1.3 for instructions for connecting these gas sources.

While it is possible to perform this procedure with any range setting we recommend that you perform this initial checkout using the 500 ppm range.

NOTE

The following procedure assumes that the instrument does not have any of the available

Zero/Span Valve Options installed.

See Section 7.4 for instructions for calibrating instruments possessing Z/S valve options.

1. Set the Analog Output Range:

Press this button to set the analyzer for SNGL DUAL or AUTO ranges

To change the value of the

reporting range span, enter the number by pressing the key under each digit until the expected value appears.

SAMPLE

RANGE = 500.000 PPM

CO2 =X.XXX

 

< TST TST > CAL

 

 

 

 

SETUP

 

SETUP X.X

 

 

 

 

 

 

 

CFG DAS RNGE PASS CLK MORE

 

EXIT

 

SETUP X.X RANGE CONTROL MENU

 

 

 

MODE SET UNIT

 

 

 

 

EXIT

Press this button to select the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

concentration units of measure:

 

 

 

 

 

 

 

 

PPB, PPM, UGM, MGM

SETUP X.X

 

RANGE: 500.000 CONC

 

 

0

0

5

0

0

.0

ENTR

EXIT

EXIT ignores the new setting and

 

 

 

 

 

 

 

 

returns to the RANGE CONTROL

 

 

 

 

 

 

 

 

MENU.

SETUP X.X

 

RANGE: 500.000 Conc

 

ENTR accepts the new setting and

 

 

returns to the

 

 

 

 

 

 

 

 

RANGE CONTROL MENU.

0

0

0

5

0

.0

ENTR

EXIT

 

Teledyne Analytical Instruments

37