Teledyne gfc 7000ta operation manual Default Hessen Status Bit Assignments

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 176
Image 176

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.15.4.7. Setting Hessen Protocol Status Flags

Teledyne Instruments’ implementation of Hessen protocols includes a set of status bits that the instrument includes in responses to inform the host computer of its condition. Each bit can be assigned to one operational and warning message flag. The default settings for these bit/flags are:

Table 6-31: Default Hessen Status Bit Assignments

STATUS FLAG NAME

DEFAULT BIT

 

ASSIGNMENT

WARNING FLAGS

 

SAMPLE FLOW WARNING

0001

 

 

BENCH TEMP WARNING

0002

SOURCE WARNING

0004

 

 

BOX TEMP WARNING

0008

 

 

WHEEL TEMP WARNING

0010

SAMPLE TEMP WARNING

0020

 

 

SAMPLE PRESSURE WARNING

0040

INVALID CONC

0080

(The Instrument’s Front Panel Display Will Show The

 

Concentration As “XXXX”)

 

OPERATIONAL FLAGS

 

Instrument Off

0100

 

 

In Manual Calibration Mode

0200

 

 

In Zero Calibration Mode

0400

In Span Calibration Mode

0800

 

 

UNITS OF MEASURE FLAGS

 

UGM

0000

 

 

MGM

2000

 

 

PPB

4000

PPM

6000

 

 

SPARE/UNUSED BITS

100, 1000, 8000

UNASSIGNED FLAGS (0000)

Sync Warning

Relay Board Warning

 

 

Conc Alarm 1

Analog Cal Warning

Conc Alarm 2

Cannot Dyn Zero

 

 

Photo Temp Warning

Cannot Dyn Span

System Reset

Invalid Conc

 

 

Rear Board Not Detected

 

 

 

NOTES:

It is possible to assign more than one flag to the same Hessen status bit. This allows the grouping of similar flags, such as all temperature warnings, under the same status bit.

Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the assigned flags is active.

Teledyne Analytical Instruments

158

Page 176
Image 176
Teledyne gfc 7000ta operation manual Default Hessen Status Bit Assignments, Status Flag Name Default BIT Assignment