10.3.4. Inability to Zero – Touchscreen ZERO Button Not Visible

1.Confirm that there is a good source of zero air. Dilute a tank of span gas with the same amount of zero air from two different sources. If the CO2 Concentration of the two measurements is different, there is a problem with one of the sources of zero air.

2.Check for leaks in the pneumatic systems as described in Section 9.3.3.

3.If the analyzer has had zero/span valve options 52 or 53, the CO2 scrubber may need maintenance.

4.Check to make sure that there is no ambient air leaking into zero air line.