Teledyne gfc 7000ta Other Performance Problems, Temperature Problems, Box or Sample Temperature

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 256
Image 256

Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

10.4. Other Performance Problems

Dynamic problems (i.e. problems which only manifest themselves when the analyzer is monitoring sample gas) can be the most difficult and time consuming to isolate and resolve. The following provides an itemized list of the most common dynamic problems with recommended troubleshooting checks and corrective actions.

10.4.1. Temperature Problems

Individual control loops are used to maintain the set point of the absorption bench, filter wheel, and IR photo-detector temperatures. If any of these temperatures are out of range or are poorly controlled, the GFC 7000TA will perform poorly.

10.4.1.1. Box or Sample Temperature

Box Temperature

The box temperature sensor is mounted to the motherboard and cannot be disconnected to check its resistance. Rather check the BOX TEMP signal using the SIGNAL I/O function under the DIAG Menu (See Section 10.1.3). This parameter will vary with ambient temperature, but at ~30oC (6-7° above room temperature) the signal should be ~1450 mV.

Sample Temperature

The Sample Temperature should closely track the bench temperature. If it does not, locate the sensor, which is located at the midpoint of the optical bench in a brass fitting. Unplug the connector labeled “Sample”, and measure the resistance of the thermistor; at room temperature (25°C) it should be ~30K Ohms, at operating temperature, 48°C, it should be ~ 12K Ohms

10.4.1.2. Bench Temperature

There are three possible failures that could cause the Bench temperature to be incorrect.

1.The heater mounted to the bottom of the Absorption bench is electrically shorted or open. Check the resistance of the two heater elements by measuring between pin 2 and 4 (~76 Ohms), and pin 3 and 4 (~330 Ohms), of the white five-pin connector just below the sample temperature sensor on the Bench (pin 1 is the pointed end).

2.Assuming that the I2C bus is working and that there is no other failure with the relay board, the solid-state relay (K2) on the relay board may have failed. Using the BENCH_HEATER parameter under the signal I/O function, as described above, turn on and off K2 (D3 on the relay board should illuminate as the heater is turned on). Check the AC voltage present between pin 2 and 4, for a 100 or 115 VAC model, and pins 3 and 4, for a 220-240 VAC model.

WARNING:

HAZARDOUS VOLTAGES ARE PRESENT DURING THIS TEST

Teledyne Analytical Instruments

238

Page 256
Image 256
Teledyne gfc 7000ta Other Performance Problems, Temperature Problems, Box or Sample Temperature, Bench Temperature