Teledyne gfc 7000ta operation manual Basic Pneumatic Connections, Rear Panel Pneumatic Connections

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 44
Image 44

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

3.6.1. Basic Pneumatic Connections

NOTE

In order to prevent dust from getting into the gas flow channels of your analyzer, it was shipped with small plugs inserted into each of the pneumatic fittings on the back panel. Remove these plugs before operating the analyzer. It is recommended that they be stored for future use (moving, storing or shipping the analyzer).

Figure 3-4 illustrates the basic configuration for gas supply and exhaust lines to the Model GFC 7000TA Analyzer. Figure 3-5 illustrates the internal gas flow of the instrument in its basic configuration.

Please refer to Figure 3-2 for pneumatic connections at the rear panel and Table 3-2 for description.

NOTE

Sample and calibration gases should only come into contact with PTFE (Teflon),

FEP, glass, stainless steel or brass.

 

Table 3-7: Rear Panel Pneumatic Connections

 

 

 

REAR PANEL LABEL

 

FUNCTION

 

 

Connect a gas line from the source of sample gas here.

SAMPLE

 

Calibration gasses are also inlet here on units without zero/span valve or IZS

 

 

options installed.

EXHAUST

 

Connect an exhaust gas line of not more than 10 meters long here.

PRESSURE SPAN

 

On units with zero/span valve options installed, connect a gas line to the source

 

of calibrated span gas here.

 

 

VENT SPAN

 

Span gas vent outlet for units with zero/span valve options installed.

 

Connect an exhaust gas line of not more than 10 meters long here.

 

 

IZS

 

Internal zero air scrubber.

 

On units with zero/span valve options installed but no internal zero air

 

 

scrubber, attach a gas line to the source of zero air here.

PURGE IN

 

This inlet supplies purge air to the GFC wheel housing (see Section 10.2.3)

 

Connect a source of dried air that has been scrubbed of CO2.

 

 

Teledyne Analytical Instruments

26

Page 44
Image 44
Teledyne gfc 7000ta operation manual Basic Pneumatic Connections, Rear Panel Pneumatic Connections