Teledyne gfc 7000ta operation manual Internal Electronic Status LEDs, CPU Status Indicator

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 242
Image 242

Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

 

PRIMARY SETUP MENU

 

 

CFG DAS RNGE PASS CLK MORE

 

EXIT

 

 

 

 

 

 

 

 

SETUP X.X SECONDARY SETUP MENU

 

COMM

VARS

DIAG

 

EXIT

 

 

 

 

 

 

 

 

SAMPLE

ENTER SETUP PASS : 818

 

 

8

 

1

8

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

 

DIAG

 

SIGNAL I/O

 

 

PREV

NEXT

ENTR

EXIT

 

 

 

 

 

 

 

 

DIAG I/O

 

0 ) EXT_ZERO_CAL=ON

 

PREV NEXT JUMP

PRNT EXIT

 

 

 

 

 

 

 

If parameter is an

 

input signal

 

DIAG I/O 28) SAMPLE_PRESSURE=2540 MV

PREV NEXT JUMP

PRNT EXIT

 

 

If parameter is an output

signal or control

DIAG I/O

22) WHEEL_HTR=ON

PREV NEXT JUMP

ON PRNT EXIT

 

 

 

Toggles parameter

ON/OFF

DIAG I/O

22) WHEEL_HTR=OFF

PREV NEXT JUMP

OFF PRNT EXIT

 

 

 

Exit returns to

DIAG display & all values return to software control

Figure 10-2: Example of Signal I/O Function

10.1.4. Internal Electronic Status LEDs

Several LED’s are located inside the instrument to assist in determining if the analyzer’s CPU, I2C bus and relay board, GFC wheel and the sync/demodulator board are functioning properly.

10.1.4.1. CPU Status Indicator

DS5, a red LED, that is located on upper portion of the motherboard, just to the right of the CPU board, flashes when the CPU is running the main program loop. After power-up, approximately 30 to 60

Teledyne Analytical Instruments

224

Page 242
Image 242
Teledyne gfc 7000ta operation manual Internal Electronic Status LEDs, CPU Status Indicator