Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.7.2.8. RS-232 Report Function

The GFC 7000TA DAS can automatically report data to the communications ports, where they can be captured with a terminal emulation program or simply viewed by the user.

To enable automatic COM port reporting, follow the instruction shown in Section 6.7.2.2 then press:

From the DATA ACQUISITION menu

(see Section 6.7.2.2)

Edit Data Channel Menu

SETUP X.X

0) CONC:

ATIMER,

1,

900

PREV NEXT

INS

DEL EDIT

PRNT

EXIT

 

 

 

 

 

 

 

 

SETUP X.X

NAME:CONC

 

 

<SET SET> EDIT PRINT

 

EXIT

 

 

 

 

 

Exits to the main Data Acquisition menu

Toggle to turn

reporting ON or OFF

Press SET> key until…

 

SETUP X.X RS-232 REPORT: OFF

 

 

<SET SET> EDIT PRINT

EXIT

 

 

 

 

 

 

 

 

 

SETUP X.X

RS-232 REPORT: OFF

 

 

OFF

 

ENTR EXIT

 

 

 

 

 

 

ENTR accepts the new setting and returns to the previous menu.

EXIT ignores the new setting and returns to the previous menu.

6.7.2.9. Compact Report

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each parameter in one channel on a separate line, up to five parameters are reported in one line.

6.7.2.10. Starting Date

This option allows to specify a starting date for any given channel in case the user wants to start data acquisition only after a certain time and date. If the Starting Date is in the past, the DAS ignores this setting.