Theory of Operation

Model GFC7000TA Carbon Dioxide Analyzer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M/R

is Shifted

IR shining through both cells is effected equally by interfering gas in the Sample Chamber

Figure 9-5: Effects of Interfering Gas on CO2 MEAS & CO2 REF

Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO2 and not to interfering gases. In this way, Gas filter correlation rejects the effects of interfering gases and so that the analyzer responds only to the presence of CO2.

To improve the signal-to-noise performance of the IR photo-detector, the GFC wheel also incorporates an optical mask that chops the IR beam into alternating pulses of light and dark at six times the frequency of the measure/reference signal. This limits the detection bandwidth helping to reject interfering signals from outside this bandwidth improving the signal to noise ratio.

The IR Signal as the Photo-Detector sees it after being chopped by the GFC Wheel Screen

CO2 MEAS

CO2 REF

Figure 9-6: Chopped IR Signal

Teledyne Analytical Instruments

191

Page 209
Image 209
Teledyne gfc 7000ta operation manual Is Shifted, Effects of Interfering Gas on CO2 Meas & CO2 REF

gfc 7000ta specifications

The Teledyne GFC 7000TA is a cutting-edge, high-performance gas flow calibrator designed for both laboratory and field applications. This instrument is renowned for its accuracy, versatility, and reliability, making it an ideal choice for various industries, including aerospace, environmental monitoring, and manufacturing.

One of the standout features of the GFC 7000TA is its advanced measurement technology. It utilizes state-of-the-art thermal mass flow meters, which allow for precise fluid measurement across a wide range of flow rates. This technology ensures highly accurate readings, which are essential for applications requiring stringent compliance with regulatory standards.

The GFC 7000TA is capable of accommodating various types of gases, making it particularly versatile. Users can calibrate flow for gases such as air, nitrogen, or carbon dioxide, providing flexibility in a multitude of applications. The device also supports dual-channel capabilities, allowing simultaneous measurement and calibration of two different gas types, enhancing productivity and efficiency in processes.

Another notable characteristic of the GFC 7000TA is its user-friendly interface. Equipped with a high-resolution touchscreen display, the device offers intuitive navigation and data visualization. Users can easily set parameters, view real-time measurements, and generate printed reports. This interface reduces the learning curve for new operators, allowing organizations to implement the GFC 7000TA with minimal training.

In addition to its advanced measuring capabilities, the GFC 7000TA features a robust design for durability in harsh environments. The instrument is compliant with various international standards, ensuring its reliability and performance in various conditions. Its portable form factor makes it convenient for fieldwork, while the rugged construction provides peace of mind for users working in demanding settings.

The GFC 7000TA also incorporates modern connectivity options. With built-in USB and Ethernet ports, users can easily integrate the device into their existing systems for data logging and remote monitoring. This connectivity feature enables real-time data access and enhances the potential for analysis and optimization of gas flow systems.

In summary, the Teledyne GFC 7000TA is a high-performance gas flow calibrator characterized by its advanced thermal mass flow measurement technology, versatility in accommodating various gases, user-friendly interface, and robust design. With its modern connectivity features, this instrument stands out as a vital tool for professionals seeking precision and reliability in gas flow applications.