Model GFC7000TA Carbon Dioxide Analyzer

6.12. SETUP MORE VARS: Internal Variables (VARS)

The GFC 7000TA has several-user adjustable software variables, which define certain operational parameters. Usually, these variables are automatically set by the instrument’s firmware, but can be manually re-defined using the VARS menu. Table 6-14 lists all variables that are available within the 818 password protected level.

 

Table 6-14: Variable Names (VARS) Revision B.3

 

 

 

 

 

NO.

VARIABLE

DESCRIPTION

ALLOWED

VALUES

 

 

 

 

 

 

 

 

 

Changes the internal data acquisition system (DAS) hold-off

 

 

 

time, which is the duration when data are not stored in the

Can be between 0.5

 

 

DAS because the software considers the data to be

0

DAS_HOLD_OFF

and 20 minutes

questionable. That is the case during warm-up or just after

 

 

 

the instrument returns from one of its calibration modes to

Default=15 min.

 

 

SAMPLE mode. DAS_HOLD_OFF can be disabled entirely

 

 

 

 

 

in each DAS channel.

 

 

 

 

 

1

CONC_PRECISION

Allows the user to set the number of significant digits to the

AUTO, 1, 2, 3, 4

right of the decimal point display of concentration and

Default=AUTO

 

 

stability values.

 

 

 

 

 

 

 

2

STABIL_GAS

Selects which gas measurement is displayed when the STABIL

CO2 & O21

 

 

test function is selected.

 

3

DYN_ZERO

Dynamic zero automatically adjusts offset and slope of the

 

CO2 response when performing a zero point calibration

ON/OFF

 

 

during an AutoCal (Section 7).

 

 

 

 

 

 

 

Dynamic span automatically adjusts slope and slope of the

 

 

 

CO2 response when performing a zero point calibration

 

4

DYN_SPAN

during an AutoCal (Section 7).

ON/OFF

 

 

 

Note that the DYN_ZERO and DYN_SPAN features are not

 

 

 

allowed for applications requiring EPA equivalency.

 

 

 

 

 

5

CLOCK_ADJ

Adjusts the speed of the analyzer’s clock. Choose the +

 

sign if the clock is too slow, choose the - sign if the clock is

-60 to +60 s/day

 

 

too fast.

 

1O2 option is only available in analyzers with O2 sensor options installed.