Model GFC7000TA Carbon Dioxide Analyzer

6.13.9. Test Channel Output

When activated, output channel A4 can be used to report one of the test functions viewable from the SAMPLE mode display. To activate the A4 channel and select a test function, follow this button sequence:

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

SETUP X.X PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X SECONDARY SETUP MENU

 

COMM VARS DIAG ALRM

EXIT

SETUP X.X

ENTER DIAG PASS: 818

8 1 8

ENTR EXIT

DIAG

SIGNAL I / O

NEXT

ENTR EXIT

DIAG

ANALOG OUTPUT

PREV NEXT

ENTR EXIT

Continue to press NEXT until …

DIAG

 

TEST CHAN OUTPUT

 

PREV

NEXT

ENTR

EXIT

DIAG TCHN

TEST CHANNEL: NONE

 

 

NEXT

ENTR

EXIT

DIAG TCHN TEST CHANNEL: CO2 MEASURE

PREV

NEXT

ENTR

EXIT

Press PREV or NEXT

Press ENTR to select

Press EXIT to

to move through the

the displayed

list of available

 

parameter activating

return to the

parameters

 

the test channel.

DIAG menu

(Table 6-13)

Table 6-22: Test Parameters Available for Analog Output A4

TEST CHANNEL

ZERO

FULL SCALE

 

 

 

NONE

Test Channel is turned off

 

 

 

CO2 MEASURE

0 mV

5000 mV*

 

 

 

CO2 REFERENCE

0 mV

5000 mV*

 

 

 

SAMPLE PRESS

0 "Hg

40 "Hg

 

 

 

SAMPLE FLOW

0 cc/m

1000 cc/m

 

 

 

SAMPLE TEMP

0°C

70°C

 

 

 

BENCH TEMP

0°C

70°C

 

 

 

WHEEL TEMP

0°C

70°C

 

 

 

CHASSIS TEMP

0°C

70°C

 

 

 

PHT DRIVE

0 mV

5000 mV

Teledyne Analytical Instruments

140