Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.9. SETUP PASS: Password Feature

The GFC 7000TA provides password protection of the calibration and setup functions to prevent unauthorized adjustments. When the passwords have been enabled in the PASS menu item, the system will prompt the user for a password anytime a password-protected function is requested.

There are three levels of password protection, which correspond to operator, maintenance, and configuration functions. Each level allows access to all of the functions in the previous level.

Table 6-9: Password Levels

PASSWORD

LEVEL

MENU ACCESS ALLOWED

No password

Operator

TEST, MSG, CLR

 

 

 

818

Maintenance

CAL, CALZ, CALS

 

 

 

929

Configuration

SETUP, VARS, DIAG

To enable or disable passwords, press the following keystroke sequence: