Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.15.2.6. Remote Access by Modem

The GFC 7000TA can be connected to a modem for remote access. This requires a cable between the analyzer’s COM port and the modem, typically a DB-9F to DB-25M cable (available from Teledyne Instruments with part number WR0000024).

Once the cable has been connected, check to make sure the DTE-DCE is in the correct position. Also make sure the GFC 7000TA COM port is set for a baud rate that is compatible with the modem, which needs to operate with an 8-bit word length with one stop bit.

The first step is to turn on the MODEM ENABLE communication mode (Mode 64, Section 6.11.5). Once this is completed, the appropriate setup command line for your modem can be entered into the analyzer. The default setting for this feature is

AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0

This string can be altered to match your modem’s initialization and can be up to 100 characters long.

To change this setting press:

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

EXIT returns to the previous menu

SETUP X.X PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

X.X SECONDARY SETUP MENU

 

COMM VARS DIAG ALRM

EXIT

SETUP X.X

COM1 MODE:0

 

SET>

EDIT

EXIT

 

 

 

 

 

 

SETUP X.X

COM1 BAUD RATE:19200

 

<SET SET>

EDIT

EXIT

 

 

 

SETUP X.X

COMMUNICATIONS MENU

 

ID INET COM1

COM2

EXIT

 

 

 

SETUP X.X

COM1 MODEM INIT:AT Y&D&H

<SET SET>

EDIT

EXIT

 

 

 

 

SETUP X.X

COM1 MODEM INIT:[A]T Y&D&H

<CH CH>

INS DEL [A]

ENTR EXIT

 

 

 

ENTR accepts the new string and returns to the previous menu. EXIT ignores the new string and returns to the previous menu.

<CH and CH> move the [ ] cursor left and right along the text string

 

 

Press the [?]

 

 

key repeatedly to cycle through the

INS inserts a

 

available character set:

DEL deletes a

0-9

character before

character at the

A-Z

the cursor location.

cursor location.

space ’ ~ ! © # $ % ^ & * ( ) - _ =

 

 

 

 

+[ ] { } < >\ ; : , . / ?

Teledyne Analytical Instruments

150

Page 168
Image 168
Teledyne gfc 7000ta operation manual Teledyne Analytical Instruments 150