Model GFC7000TA Carbon Dioxide Analyzer

In order to compensate for CPU clocks which run faster or slower, you can adjust a variable called CLOCK_ADJ to speed up or slow down the clock by a fixed amount every day. To change this variable, press:

SAMPLE

RANGE = 500.000 PPM

CO2 =X.XXX

 

< TST TST > CAL

 

SETUP

 

 

 

 

 

 

 

 

 

SETUP X.X PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

 

SETUP X.X SECONDARY SETUP MENU

 

COMM VARS DIAG

 

EXIT

 

 

 

 

 

 

 

 

 

SETUP X.

0 ) DAS_HOLD_OFF=15.0 Minutes

 

NEXT JUMP

EDIT PRNT EXIT

 

 

 

 

 

 

 

 

 

 

 

SETUPX.X 1 ) CONC_PRECISION = 3

PREV NEXT JUMP

EDIT PRNT EXIT

Continue to press NEXT until …

SETUP X.X 4) CLOCK_ADJ=0 Sec/Day

 

 

PREV

 

JUMP

EDIT PRNT EXIT

 

 

 

 

 

 

 

 

 

 

SETUP

.X

CLOCK_ADJ:0 Sec/Day

 

 

+ 0

0

 

ENTR EXIT

 

 

 

 

 

 

 

 

Enter sign and number of seconds per day the clock gains (-) or loses (+).

SETUP X.X

4) CLOCK_ADJ=0 Sec/Day

PREV NEXT JUMP

EDIT PRNT EXIT

 

 

 

 

 

3x EXIT returns

 

 

to the main SAMPLE display