Teledyne gfc 7000ta Sensor Assembly, Sync/Demodulator Assembly, Relay Board Control Devices

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 261
Image 261

Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

10.5.6. Relay Board

The relay board PCA (04135) can be most easily checked by observing the condition of the its status LEDs on the relay board, as described in Section 10.1.4.3, and the associated output when toggled on and off through signal I/O function in the diagnostic menu, See Section 10.1.3.

1.If the front panel display responds to button presses and D1 on the relay board is not flashing, then either the I2C connection between the motherboard and the relay board is bad, or the relay board itself is bad.

2.If D1 on the relay board is flashing and the status indicator for the output in question (heater power, valve drive, etc.) toggles properly using the signal I/O function, then the associated control device on the relay board is bad. Several of the control devices are in sockets and can be easily replaced. The table below lists the control device associated with a particular function:

Table 10-8: Relay Board Control Devices

FUNCTION

CONTROL

IN SOCKET

DEVICE

 

 

Wheel Heater

K1

Yes

 

 

 

Bench Heater

K2

Yes

Spare AC Control

K3

Yes

IZS Valves

U4

Yes

 

 

 

IR Source Drive

U5

No

 

 

 

The IR source drive output can be verified by measuring the voltage at J16 with the IR source disconnected. It should be 11.5± 0.5 VDC.

10.5.7. Sensor Assembly

10.5.7.1. Sync/Demodulator Assembly

To verify that the Sync/Demodulator Assembly is working follow the procedure below:

1.Verify that D1 and D2 are flashing (they flash at different rates, see Table 11-3).

If not check the opto pickup assembly, Section 10.5.7.2 and the GFC wheel drive, Section 11.5.6.3.

If the wheel drive and opto pickup are working properly then verify that there is 2.4 ±0.1 VAC and 2.5 ±0.15 VDC between digital ground and TP 5 on the sync demod board. If not then check the wiring between the sync/demod and opto pickup assembly (see interconnect drawing 04216). If good then the sync/demod board is bad.

2.Verify that the IR source is operating, Section 10.5.7.4.

3.With the analyzer connected to zero air, measure between TP11 (measure) and analog ground, and TP12 (reference) and analog ground.

If they are similar to values recorded on the factory data sheet then there is likely a problem with the wiring or the A/D converter.

Teledyne Analytical Instruments

243

Page 261
Image 261
Teledyne gfc 7000ta operation manual Sensor Assembly, Sync/Demodulator Assembly, Relay Board Control Devices