Model GFC7000TA Carbon Dioxide Analyzer

To begin the Analog Output Step Test press:

SAMPLE

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

SETUP X.X SECONDARY SETUP MENU

COMM VARS DIAG ALRM

EXIT

 

 

 

SETUP X.X

ENTER DIAG PASS: 818

8 1 8

ENTR EXIT

 

 

 

DIAG

SIGNAL I / O

 

 

NEXT

ENTR

EXIT

 

 

 

 

 

 

 

 

 

DIAG

ANALOG OUTPUT

 

 

PREV NEXT

ENTR

EXIT

 

 

 

 

 

 

 

 

 

DIAG AOUT

ANALOG OUTPUT

 

 

0%

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

DIAG AOUT

ANALOG OUTPUT

 

 

[0%]

 

EXIT

 

 

 

 

 

 

 

 

Pressing the “0%” button while performing the test will pause the test at that level. Brackets will appear around the value: example: [20%] Pressing the same button again will resume the test.

Performs

analog output

step test.

0% - 100%

Exit-Exit

returns to the

DIAG menu

6.13.4. Analog I/O Configuration

The analog I/O functions that are available in the GFC 7000TA are:

 

Table 6-16: DIAG - Analog I/O Functions

 

 

Sub Menu

Function

 

 

AOUTS CALIBRATED:

Shows the status of the analog output calibration (YES/NO) and initiates a calibration of all

 

analog output channels.

CONC_OUT_1

Sets the basic electronic configuration of the A1 analog output (CO2). There are three options:

 

Range: Selects the signal type (voltage or current loop) and full scale level of the output.

 

REC_OFS: Allows setting a voltage offset (not available when RANGE is set to CURRent loop.

 

Auto_CAL: Performs the same calibration as AOUT CALIBRATED, but on this one channel only.

 

NOTE: Any change to RANGE or REC_OFS requires recalibration of this output.

 

 

CONC_OUT_2

Same as for CONC_OUT_1 but for analog channel 2 (CO2 )

TEST OUTPUT

Same as for CONC_OUT_1 but for analog channel 4 (TEST)

AIN CALIBRATED

Shows the calibration status (YES/NO) and initiates a calibration of the analog to digital

 

converter circuit on the motherboard.

XIN1

For each of 8 external analog input channels, shows the gain, offset, engineering units, and

.

whether the channel is to show up as a Test function.

 

.

 

.

 

XIN8

 