Teledyne gfc 7000ta operation manual Step Three Perform the Zero/Span Calibration Procedure

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 182
Image 182

Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

STEP THREE: Perform the Zero/Span Calibration Procedure:

SAMPLE*

RANGE = 500.000 PPM

CO2 =XXX.X

< TST TST > CAL

SETUP

 

 

 

 

 

 

SAMPLE

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST > CAL

SETUP

 

 

 

Set the Display to show the STABIL test function.

This function calculates the

stability of the CO2 measurement

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST > CAL

SETUP

 

 

 

 

 

 

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST > ZERO CONC

EXIT

 

 

 

Wait until STABIL falls below 1.0 ppm.

This may take several

minutes.

The SPAN button

appears.

If either the ZERO or SPAN buttons fail to appear see Section 11 for troubleshooting tips.

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

Press ENTR to changes the

OFFSET & SLOPE values for the

 

 

 

 

 

< TST TST > ENTR

CONC

EXIT

CO2 measurements.

Press EXIT to leave the calibration

 

 

 

 

 

unchanged and return to the

 

 

 

 

 

 

 

 

 

 

previous menu.

 

 

ACTION:

 

 

Allow span gas to enter the sample port at the

 

 

rear of the instrument.

 

 

 

 

 

 

 

The value of

 

 

 

 

 

STABIL may jump

 

 

 

 

 

significantly.

 

 

 

 

 

Wait until it falls back

 

 

 

 

 

below 1.0 ppm

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

This may take several

 

 

 

 

 

minutes.

< TST TST >

SPAN

CONC

EXIT

 

 

 

 

 

 

Press ENTR to change the

 

 

 

 

 

M-P CAL

 

= 500.000 PPM

CO2 =XXX.X

OFFSET & SLOPE values for the

 

 

 

 

 

CO2 measurements.

< TST TST > ENTR

SPAN

CONC

EXIT

Press EXIT to leave the calibration

unchanged and return to the

 

 

 

 

 

 

 

 

 

 

previous menu.

 

 

 

 

M-P CAL

RANGE = 500.000 PPM

CO2 =XXX.X

 

 

 

 

 

 

EXIT returns to the main

< TST TST > ENTR

CONC

EXIT

SAMPLE display

 

 

 

 

 

 

If the ZERO or SPAN buttons are not displayed, this means that the measurement made during that part of the procedure is too far out of the allowable range to do allow a reliable calibration. The reason for this must be determined before the analyzer can be calibrated. See Section 11 for troubleshooting tips.

Teledyne Analytical Instruments

164

Page 182
Image 182
Teledyne gfc 7000ta operation manual Step Three Perform the Zero/Span Calibration Procedure