Teledyne gfc 7000ta operation manual Functional Check, Setup

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 54
Image 54

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

3.7.4. Functional Check

1.After the analyzer’s components has warmed up for at least 30 minutes, verify that the software properly supports any hardware options that were installed.

2.Check to make sure that the analyzer is functioning within allowable operating parameters. Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as their expected values. These functions are also useful tools for diagnosing performance problems with your analyzer (Section11.1.2). The enclosed Final Test and Validation Data sheet (part number 04307) lists these values before the instrument left the factory.

To view the current values of these parameters press the following button sequence on the analyzer’s front panel. Remember until the unit has completed its warm up these parameters may not have stabilized.

SAMPLE

RANGE = 500.000 PPM

CO2 = XXX.X

 

 

< TST TST > CAL

SETUP

 

 

 

 

RANGE

 

 

 

 

 

 

 

 

 

 

 

 

RANGE11

 

 

Toggle <TST TST> buttons

RANGE21

 

 

O2 RANGE2

 

 

to scroll through list of

 

 

 

 

 

 

STABIL

 

 

 

 

 

 

MEAS

 

 

 

 

 

 

REF

 

 

 

 

 

 

MR RATIO

Refer to

 

 

 

 

PRES

 

 

 

 

Table 6-2 for

 

 

 

 

SAMP FL

 

 

 

 

definitions of

 

 

 

 

SAMP TEMP

 

 

 

 

these test

 

 

 

 

BENCH TEMP

functions.

 

 

 

 

WHEEL TEMP

 

 

 

 

 

 

BOX TEMP

 

 

1 Only appears instrument is set

PHT DRIVE

 

 

for DUAL or AUTO reporting

SLOPE

 

 

range modes.

OFFSET

 

 

2 Only appears if O2 Sensor

TEST

 

 

Option is installed.

TIME

 

 

If your network is running a dynamic host configuration protocol (DHCP) software package, the Ethernet feature will automatically configure its interface with your LAN. (See Section 6.11.6.2). This configuration is useful for quickly getting an instrument up and running on a network. However, for permanent Ethernet connections, a static IP address should be used. (See Section 6.11.6.3).

Teledyne Analytical Instruments

36