Teledyne gfc 7000ta operation manual Connecting the Control Inputs, Control Input Pin-outs

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 42
Image 42

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

3.5.5. Connecting the Control Inputs

If you wish to use the analyzer to remotely activate the zero and span calibration modes, several digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.

There are two methods for energizing the control inputs. The internal +5V available from the pin labeled “+” is the most convenient method. However, if full isolation is required, an external 5 VDC power supply should be used.

CONTROL IN

A B C D E F

 

U +

 

 

ZS E P R A O N

Local Power Connections

CONTROL IN

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A B C D E F

 

U +

 

 

ZS E P R A O N

-

5 VDC Power

+

 

Supply

 

External Power Connections

The pin assignments for the digital control inputs can be found in the table below:

Table 3-6: Control Input Pin-outs

INPUT #

STATUS

ON CONDITION

DEFINITION

 

 

A

REMOTE ZERO

The Analyzer is placed in Zero Calibration mode. The mode field of

CAL

the display will read ZERO CAL R.

 

B

REMOTE

The Analyzer is placed in Span Calibration mode. The mode field of

SPAN CAL

the display will read SPAN CAL R.

 

CSPARE

DSPARE

ESPARE

FSPARE

 

 

Digital Ground

May be connected to the ground of the data logger/recorder.

 

 

 

 

 

 

U

Pull-up supply for

Input pin for +5 VDC required to activate pins A – F. This can be from

an external source or from the “+” pin of the instruments STATUS

inputs

 

 

connector.

 

 

 

+Internal +5V Internal source of +5V which can be used to actuate control inputs

Supply when connected to the U pin.