Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.13.4.5. AIN Calibration

This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary after major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG I/O CONFIGURATION MENU (See Section 6.13.1), then press:

STARTING FROM ANALOG I / O CONFIGURATION MENU

Instrument calibrates automatically

DIAG

 

ANALOG I / O CONFIGURATION

PREV

NEXT

 

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Continue pressing SET? until …

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

AIN CALIBRATED: NO

 

 

 

< SET SET>

CAL

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO CALIBRATING A/D ZERO

CALIBRATING A/D SPAN

DIAG AIO

AIN CALIBRATED: YES

 

 

 

 

 

 

 

< SET SET>

CAL

EXIT

 

 

 

 

 

 

 

Exit at any time to return to the main

DIAG menu

Exit to return to the

ANALOG I/O

CONFIGURATION

MENU

6.13.4.6. Configuring Analog Inputs (Option) Channels

To configure the analyzer’s external analog inputs option, define for each channel:

gain (number of units represented by 1 volt)

offset (volts)

engineering units to be represented in volts (each press of the touchscreen button scrolls the list of alphanumeric characters from A-Z and 0-9)

whether to display the channel in the Test functions

To access and adjust settings for the external Analog Inputs option channels press: