Teledyne gfc 7000ta Conc ATIMER, 4, Translates to the following configuration Channel No

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 96
Image 96

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.7.2.3. Editing DAS Data Channels

DAS configuration is most conveniently done through the APICOM remote control program. The following list of key strokes shows how to edit using the front panel.

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

EXIT will return to the previous SAMPLE display.

Moves the

display up &

down the list of Data Channels

Inserts a new Data Channel into the list BEFORE the Channel currently being displayed

SETUP X.X

PRIMARY SETUP MENU

 

 

 

CFG

DAS RNGE PASS CLK MORE

 

EXIT

 

Main Data Acquisition Menu

 

 

 

SETUP X.X

DATA ACQUISITION

 

 

 

VIEW EDIT

 

 

 

EXIT

 

SAMPLE

ENTER SETUP PASS : 818

 

 

8

1

8

 

ENTR

EXIT

 

 

 

 

Edit Data Channel Menu

SETUP X.X

0) CONC:

ATIMER,

1,

900

Exits to the Main

 

 

 

 

 

 

PREV NEXT

INS

DEL EDIT

PRNT

EXIT

Data Acquisition

Menu

 

 

 

 

 

 

 

 

 

 

 

 

Exports the

 

 

 

 

 

 

configuration of all

 

 

 

 

 

 

data channels to

Deletes The Data

 

 

 

RS-232 interface.

 

 

 

 

Channel currently

 

 

 

 

being displayed

 

 

 

 

Moves the display

between the

PROPERTIES for this

data channel.

SETUP X.X NAME:CONC

 

<SET SET> EDIT PRNT

EXIT

Exits returns to the

previous Menu

Allows to edit the channel name, see next key sequence.

Reports the configuration of current data channels to the RS-232 ports.

When editing the data channels, the top line of the display indicates some of the configuration parameters. For example, the display line:

0)CONC: ATIMER, 4, 800

translates to the following configuration:

Channel No.: 0

NAME: CONC

TRIGGER EVENT: ATIMER

PARAMETERS: Four parameters are included in this channel

EVENT: This channel is set up to record 800 data points.

Teledyne Analytical Instruments

78

Page 96
Image 96
Teledyne gfc 7000ta operation manual Conc ATIMER, 4, Translates to the following configuration Channel No