Teledyne gfc 7000ta operation manual Dark Calibration Test, Repeat Pressing Next until

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 155
Image 155

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.13.6. Dark Calibration Test

The dark calibration test interrupts the signal path between the IR photo-detector and the remainder of the sync/demod board circuitry. This allows the instrument to compensate for any voltage levels inherent in the sync/demod circuitry that might effect the calculation of CO2 concentration. Performing this calibration returns two offset voltages, One for CO2 MEAS and on for CO2 REF that are automatically added to the CPU’s calculation routine. The two offset voltages from the last calibration procedure may be reviewed by the user via the front panel display.

To activate the dark calibration procedure or review the results of a previous calibration, press:

SAMPLE

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

SETUP X.X

SECONDARY SETUP MENU

 

COMM VARS DIAG ALRM

EXIT

 

 

 

SETUP X.X

ENTER DIAG PASS: 818

8 1 8

ENTR EXIT

 

 

Electric offset for Reference signal

DIAG DARK REF DARK OFFSET: 0.0 mV

EXIT

Electric offset for Measurement signal

DIAG DARK MEAS DARK OFFSET: 0.0 mV

EXIT

DIAG

SIGNAL I / O

 

NEXT

ENTR

EXIT

 

 

 

Repeat Pressing NEXT until . . .

DIAG

DARK CALIBRATION

 

 

 

 

 

PREV NEXT

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG DARK

CO2 DARK CALIBRATION

 

 

 

Exit returns

 

 

 

 

 

 

 

 

VIEW CAL

 

EXIT

 

 

 

to the

 

 

 

 

previous menu

 

 

 

 

 

 

 

 

 

 

 

 

Calibration runs automatically

 

 

 

 

 

 

 

 

 

 

 

Display

 

 

DIAG DARK DARK CAL 1% COMPLETE

 

 

 

 

 

 

 

 

 

 

tracks %

 

 

 

 

 

 

EXIT

 

complete

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG DARK DARK CALIBRATION ABORTED

 

 

 

 

 

 

 

 

 

 

 

 

Teledyne Analytical Instruments

137

Page 155
Image 155
Teledyne gfc 7000ta operation manual Dark Calibration Test, Repeat Pressing Next until