Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

7.3. Manual Calibration Checks

Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are recommended as a regular maintenance item and in order to monitor the analyzer’s performance. To carry out a calibration check rather than a full calibration, follow these steps.

STEP ONE: Connect the sources of zero air and span gas as shown in Figures 7.1 or 7.2.

STEP TWO: Perform the zero/span calibration check procedure:

 

 

ACTION:

 

 

 

Supply the instrument with zero gas.

 

 

 

 

 

SAMPLE

RANGE = 500.0 PPM

CO2=X.XXX

Scroll the display to the

 

 

 

STABIL test function.

< TST TST > CAL

SETUP

 

 

 

 

Wait until STABIL is below 1.0 ppm. This may take several minutes.

The value of STABIL may jump significantly.

Wait until it falls

SAMPLE

STABIL=XXX.X PPM

CO2=X.XXX

< TST TST > CAL

SETUP

 

 

 

 

 

 

SAMPLE

STABIL=XXX.X PPM

CO2=X.XXX

< TST TST > CAL

SETUP

 

 

 

ACTION:

Supply span gas to the instrument

ACTION:

Record the CO2

concentration

reading.

below 1.0 ppm. This

may take several

minutes.

 

 

 

ACTION:

SAMPLE

STABIL=XXX.X PPM

CO2=X.XXX

Record the CO2

 

 

 

concentration

< TST TST > CAL

SETUP

reading.

 

 

 

 

 

The SPAN key appears during the transition from zero to

span. You may see both SPAN and ZERO buttons.

Teledyne Analytical Instruments

165

Page 183