Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

To view the TEST Functions press:

 

 

SAMPLE

 

RANGE = 500.000 PPM

 

CO2 = XXX.X

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

< TST TST > CAL

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RANGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RANGE11

 

 

Toggle <TST TST> buttons

 

 

RANGE21

 

 

 

O2 RANGE

 

 

to scroll through list of

 

 

 

 

 

 

 

 

 

 

STABIL

 

 

 

 

 

 

 

 

 

MEAS

 

 

 

 

 

 

 

 

 

REF

 

 

 

 

 

 

 

 

MR RATIO

Refer to

 

 

 

 

 

 

 

PRES

 

 

 

 

 

 

 

Table 6-2 for

 

 

 

 

 

 

 

SAMP FL

 

 

 

 

 

 

 

definitions of

 

 

 

 

 

SAMP TEMP

 

 

 

 

 

these test

 

 

 

 

 

BENCH TEMP

functions.

 

 

 

 

 

WHEEL TEMP

 

 

 

 

 

 

 

BOX TEMP

 

 

1 Only appears instrument is set

PHT DRIVE

 

 

for DUAL or AUTO reporting

 

 

SLOPE

 

 

range modes.

 

 

 

OFFSET

 

 

2 Only appears if O2 Sensor

 

 

TEST

 

 

Option is installed.

 

 

 

TIME

 

 

Figure 6-2: Viewing TEST Functions

NOTE

A value of “XXXX” displayed for any of the TEST functions indicates an out-of-

range reading or the analyzer’s inability to calculate it.

All pressure measurements are represented in terms of absolute pressure.

Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors such as air conditioning and passing storms can cause changes in the absolute atmospheric pressure.