Model GFC7000TA Carbon Dioxide Analyzer

6.11.5. COM Port Communication Modes

Each of the analyzer’s serial ports can be configured to operate in a number of different modes, listed in Table 6-10 which can be combined by adding the mode ID numbers. For example, quiet mode, computer mode and internet-enabled mode would carry a combined mode ID of 11, the standard configuration on the GFC 7000TA COM2 port. Note that each COM port needs to be configured independently.

 

 

Table 6-10: Com Port Communication Modes

 

 

 

MODE1

ID

DESCRIPTION

QUIET

 

Quiet mode suppresses any feedback from the analyzer (DAS reports, and warning

1messages) to the remote device and is typically used when the port is communicating with a computer program such as APICOM. Such feedback is still available but a command must be issued to receive them.

COMPUTER

2

Computer mode inhibits echoing of typed characters and is used when the port is

 

communicating with a computer program, such as APICOM.

 

 

SECURITY

4

When enabled, the serial port requires a password before it will respond. The only

 

command that is active is the help screen (? CR).

 

 

HESSEN

16

The Hessen communications protocol is used in some European countries. Teledyne

PROTOCOL

Instruments part number 02252 contains more information on this protocol.

 

E, 7, 1

 

When turned on this mode switches the com port settings

 

 

from

 

2048

No parity; 8 data bits; 1 stop bit

 

 

to

 

 

Even parity; 7 data bits; 1 stop bit

RS-485

1024

Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence

 

over multidrop mode if both are enabled.

 

 

MULTIDROP

32

Multidrop protocol allows a multi-instrument configuration on a single communications

PROTOCOL

channel. Multidrop requires the use of instrument IDs.

 

ENABLE

64

Enables to send a modem initialization string at power-up. Asserts certain lines in the

MODEM

RS-232 port to enable the modem to communicate.

 

ERROR

128

Fixes certain types of parity errors at certain Hessen protocol installations.

CHECKING2

 

XON/XOFF

256

Disables XON/XOFF data flow control also known as software handshaking.

HANDSHAKE2

 

HARDWARE

8

Enables CTS/RTS style hardwired transmission handshaking. This style of data

HANDSHAKE

transmission handshaking is commonly used with modems or terminal emulation

 

 

protocols as well as by Teledyne Instrument’s APICOM software.

HARDWARE

512

Improves data transfer rate when on of the com ports.

FIFO2

 

COMMAND

4096

Enables a command prompt when in terminal mode.

PROMPT

 

 

 

1Modes are listed in the order in which they appear in the SETUP MORE com COM[1 OR 2] MODE menu

2The default sting for this feature is ON. Do not disable unless instructed to by Teledyne Instruments Customer Service personnel.

Teledyne Analytical Instruments

106