Teledyne gfc 7000ta operation manual Specifications Model GFC7000TA Carbon Dioxide Analyzer

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 23
Image 23

Specifications

Model GFC7000TA Carbon Dioxide Analyzer

2. SPECIFICATIONS AND APPROVALS

2.1. Specifications

 

Table 2-1:

Model GFC 7000TA Basic Unit Specifications

 

 

 

GFC 7000TA Parameter

 

GFC 7000TA Specification

Ranges

 

Min: 0-2 ppm Full scale

(Physical Analog Output)

 

Max: 0-2,000 ppm Full scale

 

 

Selectable, dual ranges and auto ranging supported

 

 

 

Measurement Units

 

ppb, ppm, µg/m3, mg/m3, % (user selectable)

Zero Noise

 

< 0.1 ppm (RMS)

Span Noise

 

< 1% of reading (RMS)

Lower Detectable Limit

 

< 0.2 ppm1

Zero Drift (24 hours)

 

<0.25 ppm1

Span Drift (24 hours)

 

<0.5% of reading 1

Lag Time

 

10 seconds

Rise/Fall Time

 

<60 seconds to 95%

Linearity

 

1% of full scale

Precision

 

0.5% of reading

 

 

 

Sample Flow Rate

 

800cm3/min. ±10%

 

 

O2 Sensor option adds 110 cm³/min, ±20%, to total flow through when installed.

 

 

 

Temperature Coefficient

 

< 0.1% of Full Scale per oC

Voltage Coefficient

 

< 0.05% of Full Scale per V

AC Power Rating

 

100V-120V, 220V – 240 V, 50/60 Hz

Analog Output Ranges

 

10V, 5V, 1V, 0.1V (selectable)

Analog Output Resolution

 

1 part in 4096 of selected full-scale voltage

 

 

 

Recorder Offset

 

±10%

 

 

 

Standard I/O

 

1 Ethernet: 10/100Base-T

 

 

2 RS-232 (300 – 115,200 baud)

 

 

2 USB device ports

 

 

8 opto-isolated digital status outputs

 

 

6 opto-isolated digital control inputs (3 defined, 3 spare)

 

 

4 analog outputs

 

 

 

Optional I/O

 

1 USB com port

 

 

1 RS485

 

 

8 analog inputs (0-10V, 12-bit)

 

 

4 digital alarm outputs

 

 

Multidrop RS232

 

 

3 4-20mA current outputs

 

 

 

Environmental

 

Installation category (over-voltage category) II; Pollution degree 2

 

 

 

Temperature Range

 

5-40oC

Humidity Range

 

0 - 95% RH, non-condensing

Dimensions H x W x D

 

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Teledyne Analytical Instruments

5

Page 23
Image 23
Teledyne gfc 7000ta Specifications Model GFC7000TA Carbon Dioxide Analyzer, Model GFC 7000TA Basic Unit Specifications