Page 102
Image 102

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.7.2.7. Number of Records

The number of data records in the GFC 7000TA is limited to about a cumulative one million data points in all channels (one megabyte of space on the disk-on-chip). However, the actual number of records is also limited by the total number of parameters and channels and other settings in the DAS configuration.

Every additional data channel, parameter, number of samples setting etc. and will reduce the maximum amount of data points somewhat. In general, however, the maximum data capacity is divided amongst all channels (max: 20) and parameters (max: 50 per channel).

The DAS will check the amount of available data space and prevent the user from specifying too many records at any given point. If, for example, the DAS memory space can accommodate 375 more data records, the ENTR key will disappear when trying to specify more than that number of records. This check for memory space may also make an upload of an DAS configuration with APICOM or a Terminal program fail, if the combined number of records would be exceeded. In this case, it is suggested to either try from the front panel what the maximum number of records can be or use trial-and-error in designing the DAS script or calculate the number of records using the DAS or APICOM manuals. To set the number of records for one channel from the front panel, press SETUP-DAS-EDIT-ENTR and the following touchscreen button sequence.

From the DATA ACQUISITION menu

(see Section 6.12.2.2)

Edit Data Channel Menu

SETUP X.X

0) CONC:

ATIMER,

1 2,

900

PREV NEXT

INS

DEL EDIT

PRNT

EXIT

 

 

 

 

 

 

 

 

SETUP X.X

NAME:CONC

 

 

<SET SET> EDIT PRINT

 

EXIT

 

 

 

 

 

Exits to the main Data Acquisition menu

 

 

 

Press SET> key until…

 

 

 

 

 

 

SETUP X.X

NUMBER OF RECORDS:000

 

 

<SET

SET> EDIT

PRINT

 

 

EXIT

 

 

 

 

 

 

 

SETUP X.X

EDIT RECOPRDS (DELET DATA)

YES will delete all data

YES

NO

 

 

 

 

 

in this channel.

 

 

 

 

 

 

 

 

 

 

 

 

Toggle buttons to set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

REPORT PERIODD:DAYS:0

 

number of records

 

(1-99999)

0

0

0

0

0

ENTR

EXIT

 

 

 

 

 

 

 

 

 

NO returns to the previous menu.

ENTR accepts the new setting and returns to the previous menu.

EXIT ignores the new setting and returns to the previous menu.