Image 112

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.8.5. Auto Range Mode (AUTO)

In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined ranges (low and high). The unit will switch from low range to high range when the CO2 concentration exceeds 98% of the low range span. The unit will return from high range back to low range once both the CO2 concentration falls below 75% of the low range span.

In AUTO Range mode the instrument reports the same data in the same range on both the A1 and A2 outputs and automatically switches both outputs between ranges as described above. Also, the RANGE test function displayed on the front panel will automatically switch to show which range is in effect.

The high/low range status is also reported through the external, digital status outputs (Section 6.15.1.1).

To set individual ranges press the following keystroke sequence.

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

SETUP X.X

RANGE CONTROL MENU

MODE SET UNIT

EXIT

 

 

 

 

 

 

SETUP X.X

RANGE MODE: SNGL

 

SNGL DUAL AUTO

ENTR EXIT

 

 

 

 

 

 

SETUP X.X

 

RANGE MODE: AUTO

 

SNGL DUAL AUTO

 

 

ENTR EXIT

 

 

 

 

 

 

SETUP X.X

 

RANGE CONTROL MENU

MODE SET UNIT

 

 

 

EXIT

 

 

 

 

 

 

SETUP X.X

 

LOW RANGE: 500.0 Conc

0

0

5

0

0

.0

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

SETUP X.X

 

HIGH RANGE: 500.0 Conc

0

0

5

0

0

.0

ENTR EXIT

 

 

 

 

 

 

 

EXIT x 2 returns

to the main

SAMPLE display

Toggle the numeral buttons to set the LOW and HIGH range value. ENTR accepts the new setting, EXIT ignores the new setting.

CAUTION

In AUTO range mode the LOW and HIGH ranges have separate slopes and offsets

for computing CO2 concentration.

The two ranges must be independently calibrated.

NOTE

Avoid accidentally setting the low range of the instrument with a higher span limit than the high range. This will cause the unit to stay in the low reporting range perpetually and defeat the function of the AUTO range mode.

Teledyne Analytical Instruments

94