Troubleshooting

Model GFC7000TA Carbon Dioxide Analyzer

Table 11-1: Warning Messages – Indicated Failures (cont.)

 

 

 

WARNING

FAULT CONDITION

POSSIBLE CAUSES

MESSAGE

 

 

SAMPLE TEMP

Sample temperature is < 10oC or >

Ambient temperature outside of specified range

WARN

100oC.

Failed bench heater

 

 

Failed bench temperature sensor

 

 

Relay controlling the bench heater

 

 

Failed relay board

 

 

I2C bus

SOURCE

Occurs when CO2 Ref is <1250

GFC wheel stopped

WARNING

mVDC or >4950 mVDC.

Failed sync/demod board

 

 

If status LED’s on the sync/demod board ARE flashing the

 

Either of these conditions will result

cause is most likely a failed:

 

in an invalid M/R ratio.

IR source

 

 

Relay board

 

 

I2C bus

 

 

IR photo-detector

SYSTEM RESET

The computer has rebooted.

This message occurs at power on. If you have not cycled the

 

 

power on your instrument:

 

 

o Failed +5 VDC power,

 

 

o Fatal error caused software to restart

 

 

o Loose connector/wiring

WHEEL TEMP

The filter wheel temperature is

Blocked cooling vents below GFC Assembly. Make sure that

WARNING

controlled at 68 ± 2 °C

adequate clear space beneath the analyzer.

 

 

Analyzer’s top cover removed

 

 

Wheel heater

 

 

Wheel temperature sensor

 

 

Relay controlling the wheel heater

 

 

Entire relay board

 

 

I2C bus

10.1.2. Fault Diagnosis with TEST Functions

Besides being useful as predictive diagnostic tools, the test functions viewable from the front panel can be used to isolate and identify many operational problems when combined with a thorough understanding of the analyzer’s theory of operation (see Section 10).

The acceptable ranges for these test functions are listed in the “Nominal Range” column of the analyzer Final Test and Validation Data Sheet (p/n 04307) shipped with the instrument. Values outside these acceptable ranges indicate a failure of one or more of the analyzer’s subsystems. Functions whose values are still within the acceptable range but have significantly changed from the measurement recorded on the factory data sheet may also indicate a failure. A worksheet has been provided in Appendix C to assist in recording the value of these test functions.

Table 11-2 contains some of the more common causes for these values to be out of range.