Teledyne gfc 7000ta operation manual COM Port Password Security, To initialize the modem press

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 169
Image 169

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

To initialize the modem press:

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

EXIT returns to the previous menu

SETUP X.X PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

 

X.X

SECONDARY SETUP MENU

 

COMM VARS DIAG

ALRM

EXIT

 

 

 

 

 

 

SETUP X.X

COMMUNICATIONS MENU

 

ID INET

COM1

COM2

EXIT

 

 

 

 

SETUP X.X

 

COM1 MODE:0

SET>

EDIT

EXIT

 

 

 

 

SETUP X.X

COM1 BAUD RATE:19200

<SET SET>

EDIT

EXIT

 

 

 

 

 

SETUP X.X

COM1 MODEM INIT:AT Y&D&H

<SET SET>

EDIT

EXIT

 

 

 

 

SETUP X.X

COM1 INITIALIZE MODEM

<SET SET>

INIT

EXIT

 

 

 

 

 

 

SETUP X.X

 

INITIALIZING MODEM

<SET SET>

INIT

EXIT

EXIT returns to the

Communications Menu.

6.15.2.7. COM Port Password Security

In order to provide security for remote access of the GFC 7000TA, a LOGON feature can be enabled to require a password before the instrument will accept commands. This is done by turning on the SECURITY MODE (Mode 4, Section 6.11.5). Once the SECURITY MODE is enabled, the following items apply.

A password is required before the port will respond or pass on commands.

If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the LOGOFF command.

Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be disabled for 1 hour, even if the correct password is used.

If not logged on, the only active command is the '?' request for the help screen.

The following messages will be returned at logon:

LOGON SUCCESSFUL - Correct password given

LOGON FAILED - Password not given or incorrect

LOGOFF SUCCESSFUL - Connection terminated successfully

To log on to the Model GFC 7000TA analyzer with SECURITY MODE feature enabled, type: LOGON 940331

940331 is the default password. To change the default password, use the variable RS232_PASS issued as follows:

V RS232_PASS=NNNNNN

Teledyne Analytical Instruments

151

Page 169
Image 169
Teledyne gfc 7000ta operation manual COM Port Password Security, To initialize the modem press