Teledyne gfc 7000ta operation manual From the Data Acquisition menu

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 101
Image 101

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

To define the REPORT PERIOD, follow the instruction shown in Section 6.7.2.2 then press:

From the DATA ACQUISITION menu

 

 

(see Section 6.7.2.2)

 

 

Edit Data Channel Menu

 

 

 

Use the PREV and NEXT

 

 

 

 

buttons to scroll to the

0) CONC:

ATIMER,

1,

900

data channel to be edited.

 

 

 

 

PREV NEXT

INS

DEL EDIT

PRNT

EXIT

SETUP X.X

NAME:CONC

 

 

<SET SET> EDIT PRINT

 

EXIT

Exits to the main

Data Acquisition

menu.

Press SET> key until you reach REPORT PERIOD

 

SETUP X.X

REPORT PERIOD:000:01:00

 

 

<SET

SET> EDIT

PRINT

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

Set the number of days

SETUP X.X

REPORT PERIODD:DAYS:0

 

0

0

0

 

ENTR

EXIT

between reports (0-366).

 

Press buttons to set hours

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

REPORT PERIODD:TIME:01:01

 

between reports in the format :

 

 

 

 

 

 

 

HH:MM (max: 23:59). This is a

0

1

0

0

ENTR

EXIT

24 hour clock . PM hours are 13

 

 

 

 

 

 

thru 23, midnight is 00:00.

 

 

 

 

 

 

IIf at any time an illegal entry is selected (e.g., days > 366)

Example 2:15 PM = 14:15

the ENTR button will disappear from the display.

 

ENTR accepts the new string and

returns to the previous menu. EXIT ignores the new string and returns to the previous menu.

Teledyne Analytical Instruments

83

Page 101
Image 101
Teledyne gfc 7000ta operation manual From the Data Acquisition menu