Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.11.9. COM Port Testing

The serial ports can be tested for correct connection and output in the com menu. This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is running, the red LED on the rear panel of the analyzer should flicker.

To initiate the test press the following touchscreen button sequence.

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

SETUP

 

 

 

 

SETUP X.X PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

SETUP X.X SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

 

 

 

Select which

COM port to test.

SETUP X.X

COMMUNICATIONS MENU

 

 

ID INET

COM1 COM2

EXIT

 

 

 

 

 

 

 

 

 

SETUP X.X

COM1 MODE:0

 

 

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

SETUP X.X

COM1 BAUD RATE:19200

 

 

<SET SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

SETUP X.X

COM1 : TEST PORT

 

 

<SET

TEST

EXIT

 

 

 

 

 

 

 

 

 

SETUP X.X

TRANSMITTING TO COM1

 

EXIT returns to

 

 

 

<SET

TEST

EXIT

COMM menu

 

 

 

 

 

 

 

Test runs

 

 

 

automatically

 

 

Teledyne Analytical Instruments

117