Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

To select manual output calibration for a particular channel, activate the ANALOG I/O CONFIGURATION MENU (See Section 6.13.1), then press:

Exit to return to

the main

sample display

DIAG

ANALOG I / O CONFIGURATION

 

PREV NEXT

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

AOUTS CALIBRATED: NO

 

 

< SET SET>

CAL

EXIT

 

 

 

 

Press SET> to select the analog output channel to

 

be configured. Then press EDIT to continue

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_2:5V, CAL

 

 

< SET SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_2 RANGE: 5V

 

 

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_2 REC OFS: 0 mV

< SET SET>

EDIT

EXIT

 

 

 

 

 

 

DIAG AIO CONC_OUT_2 AUTO CAL: ON

 

< SET SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_2 AUTO CAL: ON

 

ON

ENTR

EXIT

 

 

 

 

 

 

 

Toggles the

ENTR accepts the new setting

auto cal mode

and returns to the previous

ON/ OFF for

menu. EXIT ignores the new

this analog

setting and returns to the

output channel

previous menu.

only.

 

Now the analog output channels should either be automatically calibrated or they should be set to manual calibration, which is described next.

6.13.4.3. Manual Analog Output Calibration and Voltage Adjustment

For highest accuracy, the voltages of the analog outputs can be manually calibrated. Calibration is done through the instrument software with a voltmeter connected across the output terminals (Figure 6-15). Adjustments are made using the touchscreen buttons by setting the zero-point first and then the span- point (Table 6-20).

The software allows this adjustment to be made in 100, 10 or 1 count increments.

Table 6-20: Voltage Tolerances for Analog Output Calibration

Full Scale

Zero Tolerance

Span Voltage

Span Tolerance

0.1 VDC

±0.0005V

90 mV

±0.001V

1 VDC

±0.001V

900 mV

±0.001V

5 VDC

±0.002V

4500 mV

±0.003V

10 VDC

±0.004V

4500 mV

±0.006V

NOTE

Outputs configured for 0.1V full scale should always be calibrated manually

Teledyne Analytical Instruments

129

Page 147
Image 147