Model GFC7000TA Carbon Dioxide Analyzer

6.8.3. Single Range Mode (SNGL)

This is the default reporting range mode for the analyzer. In single range mode both A1 and A2 are set to the same reporting range. This reporting range can be any value between 50 ppm and 2 000 ppm.

While the two outputs always have the same reporting range, the span, signal offset and scaling of their electronic signals may be configured for differently (e.g., A1 = 0-10 V; A2 = 0-0.1 V). See Section 6.13.4 for instructions on adjusting these parameters.

To select SNGL range mode and to set the upper limit of the range, press:

SAMPLE*

RANGE = 500.000 PPM

CO2 X.XXX

< TST TST > CAL

SETUP

 

 

 

SETUP C.3

PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

SETUP C.3

RANGE MODE:

 

SNGL DUAL AUTO

ENTR EXIT

 

 

 

 

SETUP C.3

RANGE CONTROL MENU

MODE SET UNIT

 

EXIT

 

 

 

SETUP C.3

RANGE CONTROL MENU

MODE SET UNIT

 

EXIT

 

 

 

 

 

 

SETUP C.3

RANGE MODE: SNGL

 

SNGL DUAL AUTO

ENTR EXIT

 

 

 

SETUP C.3

 

RANGE: 500.0 Conc

0

0

5

0

0

.0 ENTR EXIT

SETUP C.3

 

RANGE CONTROL MENU

MODE SET UNIT

 

 

EXIT

EXIT x 2 returns

to the main

SAMPLE display