Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.8.3. Single Range Mode (SNGL)

This is the default reporting range mode for the analyzer. In single range mode both A1 and A2 are set to the same reporting range. This reporting range can be any value between 50 ppm and 2 000 ppm.

While the two outputs always have the same reporting range, the span, signal offset and scaling of their electronic signals may be configured for differently (e.g., A1 = 0-10 V; A2 = 0-0.1 V). See Section 6.13.4 for instructions on adjusting these parameters.

To select SNGL range mode and to set the upper limit of the range, press:

SAMPLE*

RANGE = 500.000 PPM

CO2 X.XXX

< TST TST > CAL

SETUP

 

 

 

SETUP C.3

PRIMARY SETUP MENU

 

CFG DAS RNGE PASS CLK MORE

EXIT

 

 

 

SETUP C.3

RANGE MODE:

 

SNGL DUAL AUTO

ENTR EXIT

 

 

 

 

SETUP C.3

RANGE CONTROL MENU

MODE SET UNIT

 

EXIT

 

 

 

SETUP C.3

RANGE CONTROL MENU

MODE SET UNIT

 

EXIT

 

 

 

 

 

 

SETUP C.3

RANGE MODE: SNGL

 

SNGL DUAL AUTO

ENTR EXIT

 

 

 

SETUP C.3

 

RANGE: 500.0 Conc

0

0

5

0

0

.0 ENTR EXIT

SETUP C.3

 

RANGE CONTROL MENU

MODE SET UNIT

 

 

EXIT

EXIT x 2 returns

to the main

SAMPLE display

Teledyne Analytical Instruments

92

Page 110
Image 110
Teledyne gfc 7000ta operation manual Single Range Mode Sngl

gfc 7000ta specifications

The Teledyne GFC 7000TA is a cutting-edge, high-performance gas flow calibrator designed for both laboratory and field applications. This instrument is renowned for its accuracy, versatility, and reliability, making it an ideal choice for various industries, including aerospace, environmental monitoring, and manufacturing.

One of the standout features of the GFC 7000TA is its advanced measurement technology. It utilizes state-of-the-art thermal mass flow meters, which allow for precise fluid measurement across a wide range of flow rates. This technology ensures highly accurate readings, which are essential for applications requiring stringent compliance with regulatory standards.

The GFC 7000TA is capable of accommodating various types of gases, making it particularly versatile. Users can calibrate flow for gases such as air, nitrogen, or carbon dioxide, providing flexibility in a multitude of applications. The device also supports dual-channel capabilities, allowing simultaneous measurement and calibration of two different gas types, enhancing productivity and efficiency in processes.

Another notable characteristic of the GFC 7000TA is its user-friendly interface. Equipped with a high-resolution touchscreen display, the device offers intuitive navigation and data visualization. Users can easily set parameters, view real-time measurements, and generate printed reports. This interface reduces the learning curve for new operators, allowing organizations to implement the GFC 7000TA with minimal training.

In addition to its advanced measuring capabilities, the GFC 7000TA features a robust design for durability in harsh environments. The instrument is compliant with various international standards, ensuring its reliability and performance in various conditions. Its portable form factor makes it convenient for fieldwork, while the rugged construction provides peace of mind for users working in demanding settings.

The GFC 7000TA also incorporates modern connectivity options. With built-in USB and Ethernet ports, users can easily integrate the device into their existing systems for data logging and remote monitoring. This connectivity feature enables real-time data access and enhances the potential for analysis and optimization of gas flow systems.

In summary, the Teledyne GFC 7000TA is a high-performance gas flow calibrator characterized by its advanced thermal mass flow measurement technology, versatility in accommodating various gases, user-friendly interface, and robust design. With its modern connectivity features, this instrument stands out as a vital tool for professionals seeking precision and reliability in gas flow applications.