Teledyne gfc 7000ta Calibration Quality, Calibration Data Quality Evaluation, Offs

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 194
Image 194

Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected

SETUP C.4

 

RANGE TO CAL: LOW

 

 

 

<SET

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP C.4

 

RANGE TO CAL: LOW

 

 

 

LOW HIGH

 

 

 

ENTR

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP C.4

 

RANGE TO CAL: HIGH

 

 

 

<SET

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP C.4

SEQ 2) ZERO–SPAN, 2:00:30

 

 

 

PREV NEXT MODE SET

EXIT

 

 

 

 

 

 

 

 

 

 

 

EXIT returns to the

PRIMARY SETUP

Menu

NOTE

In order to automatically calibrate both the HIGH and LOW ranges, you must set

up a separate sequence for each.

7.7. Calibration Quality

After completing one of the calibration procedures described above, it is important to evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear response curve of the analyzer. The values for these terms, both individually and relative to each other, indicate the quality of the calibration. To perform this quality evaluation, you will need to record the values of both test functions (Section 6.2.1 or Appendix A-3), all of which are automatically stored in the DAS channel CALDAT for data analysis, documentation and archival.

Make sure that these parameters are within the limits listed in Table 7-3 and frequently compare them to those values on the Final Test and Checkout Sheet that came attached to your manual, which should not be significantly different. If they are, refer to the troubleshooting Section 11.

Table 7-3: Calibration Data Quality Evaluation

FUNCTION

MINIMUM VALUE

OPTIMUM VALUE

MAXIMUM VALUE

SLOPE

0.700

1.000

1.300

OFFS

-0.500

0.000

0.500

These values should not be significantly different from the values recorded on the Teledyne Instruments Final Test and Validation Data sheet that was shipped with your instrument. If they are, refer to the troubleshooting Section 11.

Teledyne Analytical Instruments

176

Page 194
Image 194
Teledyne gfc 7000ta operation manual Calibration Quality, AutoCal with Auto or Dual Reporting Ranges Modes Selected, Offs