Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.13.4.1. Analog Output Signal Type and Range Span Selection

To select an output signal type (DC Voltage or current) and level for one output channel, activate the ANALOG I/O CONFIGURATION MENU (See Section 6.13.1) then press:

FROM ANALOG I/O CONFIGURATION MENU

DIAG

ANALOG I / O CONFIGURATION

 

PREV

NEXT

ENTR

EXIT

 

 

 

 

 

 

DIAG AIO

AOUTS CALIBRATED: NO

 

< SET SET>

CAL

EXIT

 

 

 

 

Press SET> to select the analog output channel to be configured. Press EDIT to continue

DIAG AIO

CONC_OUT_2:5V, CAL

 

< SET SET>

EDIT

EXIT

 

 

 

To set the signal level and type of the selected channel

DIAG AIO

 

CONC_OUT_2 RANGE: 5V

 

SET>

EDIT

EXIT

 

 

 

 

DIAG AIOOUTPUT RANGE: 5V

 

0.1V

1V

5V

10V CURR

ENTR EXIT

 

 

 

 

DIAG AIOOUTPUT RANGE: 10V

 

0.1V

1V

5V

10V CURR

ENTR EXIT

 

 

 

 

 

Pressing ENTR records the new setting and returns to the previous menu.

Pressing EXIT ignores the new setting and

returns to the previous menu.

6.13.4.2. Analog Output Calibration Mode

The analog outputs can be calibrated automatically or manually. In its default mode, the instrument is configured for automatic calibration of all channels. Manual calibration should be used for the 0.1V range or in cases where the outputs must be closely matched to the characteristics of the recording device. Outputs configured for automatic calibration can be calibrated as a group or individually. Calibration of the analog outputs needs to be carried out on first startup of the analyzer (performed in the factory as part of the configuration process) or whenever re-calibration is required.