Teledyne gfc 7000ta operation manual Step TWO Set the expected CO2 Span Gas concentration

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 181
Image 181

Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 7-2: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator

STEP TWO: Set the expected CO2 Span Gas concentration:

The CO2 span

SAMPLE*

RANGE = 500.000 PPM

CO2 =X.XXX

< TST TST > CAL

 

SETUP

 

 

 

 

 

 

 

M-P CAL

RANGE = 500. PPM

CO2 =X.XXX

< TST TST >

ZERO

CONC

EXIT

 

 

 

 

This sequence causes the analyzer to prompt for the expected CO2 span concentration.

concentration values automatically default to 400.0 Conc.

To change this value to

the actual concentration of

the span gas, enter the number by pressing the button under each digit until the expected value appears.

 

 

 

 

 

 

 

 

EXIT ignores the new setting

M-P CAL

CO2 SPAN CONC: 400.000 Conc

 

 

and returns to the previous

 

 

 

 

 

 

 

 

display.

0

0

0

4

5

.0

ENTR

EXIT

ENTR accepts the new setting

and returns to the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

previous display..

 

 

 

 

 

 

 

 

NOTE

For this Initial Calibration it is important to independently verify the PRECISE CO2 Concentration

Value of the SPAN gas.

If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration value

printed on the bottle.

Teledyne Analytical Instruments

163

Page 181
Image 181
Teledyne gfc 7000ta operation manual Step TWO Set the expected CO2 Span Gas concentration