Image 107

Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

6.8.SETUP RNGE: Analog Output Reporting Range Configuration

The analyzer has three active analog output signals, accessible through a connector on the rear panel.

ANALOG OUT

 

CO2 concentration

 

Test Channel

 

outputs

 

 

 

 

 

 

 

 

 

 

 

 

 

Not Used

 

A1

 

A2

A3

A4

 

 

+ -

+ -

+ -

+ -

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LOW range when DUAL

HIGH range when DUAL

mode is selected

mode is selected

Figure 6-6: Analog Output Connector Pin Out

All three outputs can be configured either at the factory or by the user for full scale outputs of 0.1 VDC, 1VDC, 5VDC or 10VDC. Additionally A1 and A2 may be equipped with optional 0-20 mADC current loop drivers and configured for any current output within that range (e.g. 0-20, 2-20, 4-20, etc.). The user may also adjust the signal level and scaling of the actual output voltage or current to match the input requirements of the recorder or data logger (See Section 6.13.4).

The A1 and A2 channels output a signal that is proportional to the CO2 concentration of the sample gas. Several modes are available which allow them to operate independently or be slaved together (See Section 6.7). The user may also select between a variety of reporting range spans (See Sections 6.8.3,

6.8.4and 6.8.5).

EXAMPLE:

A1 OUTPUT: Output Signal = 0-5 VDC representing 0-1000 ppm concentration values

A2 OUTPUT: Output Signal = 0 – 10 VDC representing 0-500 ppm concentration values.

Output A3 is only active if the O2 sensor option is installed. In this case a signal representing the currently measured O2 concentration is output on this channel.

The output, labeled A4 is special. It can be set by the user (See Section 6.13.9) to output several of the test functions accessible through the <TST TST> keys of the units sample display.