Teledyne gfc 7000ta operation manual Manual Calibration with Zero/Span Valves

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 184
Image 184

Calibration Procedures

Model GFC7000TA Carbon Dioxide Analyzer

7.4. Manual Calibration with Zero/Span Valves

There are four different zero/span valve option configurations (See Section 5.4). They all operate identically, differing only in the method used to supply calibration gas to the Analyzer.

STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.

Figures 7-3 through 7-6 show the proper pneumatic connections for GFC 7000TAs with various optional internal valve sets installed.

VENT

Figure 7-3: Pneumatic Connections – Ambient Zero/Pressurized Span Valves

Teledyne Analytical Instruments

166

Page 184
Image 184
Teledyne gfc 7000ta Manual Calibration with Zero/Span Valves, Pneumatic Connections Ambient Zero/Pressurized Span Valves