Image 59

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

SAMPLE

RANGE = 500.000 PPM

CO2 =XXX.X

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

< TST TST > CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

 

GAS TO CAL:CO2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO2

O2

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

 

A1:NXCNC1 =100PPM

NOX=X.XXX

 

 

 

 

 

 

 

 

 

 

 

 

 

 

<TST TST> ZERO SPAN CONC

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

 

GAS TO CAL:O2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOX

O2

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

 

O2 SPAN CONC:20.8%

 

 

 

 

EXIT ignores the new

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

2

0

.8

0

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

setting and returns to

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the previous display.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The O2 span concentration value automatically defaults to

 

 

 

 

 

 

 

 

 

 

ENTR accepts the new

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

setting and returns to

 

 

 

20.8 %.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the previous menu.

 

 

 

If this is not the the concentration of the span gas being

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

used, toggle these buttons to set the correct concentration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

of the O2 calibration gases.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STEP 2 – ACTIVATE O2 SENSOR STABILITY FUNCTION

To change the stability test function from NOx concentration to the O2 sensor output, press:

Teledyne Analytical Instruments

41

Page 59
Image 59
Teledyne gfc 7000ta operation manual Activate O2 Sensor Stability Function, Previous display