Operating Instructions

Model GFC7000TA Carbon Dioxide Analyzer

To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (See Section 6.13.1), then press:

Exit at any time to return to the main DIAG menu

STARTING FROM DIAGNOSTIC MENU

(see Section 6.13.1)

DIAG

 

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR EXIT

 

 

 

If any of the channels have not been calibrated this message will read NO.

DIAG AIO

AOUTS CALIBRATED: NO

 

< SET SET>

CAL

EXIT

DIAG AUTO CALIBRATING CONC_OUT_1

AUTO CALIBRATING CONC_OUT_2

AUTO CALIBRATING TEST_OUTPUT

If AutoCal has been turned off for any channel, the message for that channel will be similar to:

NOT AUTO CAL CONC_OUT_1

Exit to return to

DIAG AIO AOUTS CALIBRATED:

YES

< SET SET> CAL

EXIT

 

 

the I/O configuration menu

To automatically calibrate a single analog channel, activate the ANALOG I/O CONFIGURATION MENU (See Section 6.13.1), then press:

Teledyne Analytical Instruments

127

Page 145
Image 145