Teledyne gfc 7000ta operation manual Perform the Zero/Span Calibration Procedure, Stabil=Xxx.X Ppm

Models: gfc 7000ta

1 280
Download 280 pages 4.46 Kb
Page 57
Image 57

Getting Started

Model GFC7000TA Carbon Dioxide Analyzer

3. Perform the Zero/Span Calibration Procedure

SAMPLE*

RANGE = 500.000 PPM

CO2 =XXX.X

< TST TST > CAL

SETUP

 

 

 

 

 

 

SAMPLE

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST > CAL

SETUP

 

 

 

Set the Display to show the

STABIL test function.

This function calculates the

stability of the COx

measurement

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST > CAL

SETUP

 

 

 

 

 

 

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST > ZERO CONC

EXIT

 

 

 

Wait until STABIL falls below 1.0 ppm.

This may take several

minutes.

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

Press ENTR to changes the

OFFSET & SLOPE values for the

 

 

 

< TST TST > ENTR CONC

EXIT

CO2 measurements.

Press EXIT to leave the calibration

 

 

 

unchanged and return to the

 

 

 

 

 

 

previous menu.

The SPAN button now

appears during the transition from zero to span..

If either the ZERO or SPAN buttons fail to appear see Section 11 for troubleshooting tips.

ACTION:

Allow span gas to enter the sample port at the

rear of the instrument.

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST >

SPAN

CONC

EXIT

 

 

 

 

 

 

M-P CAL

= 500.000 PPM

CO2 =XXX.X

< TST TST > ENTR SPAN

CONC

EXIT

 

 

 

 

The value of

STABIL may jump

significantly.

Wait until it falls back

below 1.0 ppm

This may take several

minutes.

Press ENTR to change the

OFFSET & SLOPE values for the

CO2 measurements.

Press EXIT to leave the calibration

unchanged and return to the

previous menu.

M-P CAL

RANGE = 500.000 PPM

CO2 =XXX.X

 

 

 

 

EXIT returns to the main

< TST TST > ENTR CONC

EXIT

SAMPLE display

 

 

 

 

Teledyne Analytical Instruments

39

Page 57
Image 57
Teledyne gfc 7000ta operation manual Perform the Zero/Span Calibration Procedure, Stabil=Xxx.X Ppm