3. Preliminary Display Adjustments (SN 3001A and Below)

Procedure

X and Y DeflectionAmplifier PulseResponse Adjustments

Note

1.Connect a 10:1 (10 MQ) divider probe to the oscilloscope’s channel

1input and a 10: 1 divider probe to the channel 4 input.

2.On the oscilloscope, press CRECALL) [ml to perform a soft reset.

3.On the oscilloscope, press [CHAN] more preset probe , select channel 1, and use the front-panel knob to select a 1O:l probe.

4.Select channel 4, and use the front-panel knob to select a 1O:l probe.

5.Press @iYGiJ

6.Connect the channel 1 probe to the oscilloscope’s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press [AUTO- SCALEJ Adjust the channel 1 probe for an optimum square wave display on the oscilloscope.

7.Connect the channel 4 probe to the oscilloscope’s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press LAUTO- SCALE_). Adjust the channel 4 probe for an optimum square wave display on the oscilloscope.

Each probe is now compensated for the oscilloscope input to which it is connected. Do not interchange probes without recompensating.

8.Connect the channel 1 1O:l divider probe to AlA4E1,and the channel 4 probe to AlA4E2, as shown in Figure 3-16. Connect the probe ground leads to chassis ground. See Figure 3-17 and Figure 3-18 for the location of the assemblies and test points.

SOQ TEPMNATIC+

PULSE/FUNCTION

GENERATOR

Figure 3-16. Preliminary Display Adjustments Setup

9.Remove the cover over A3 Digital Storage Section and remove A3A2 Intensity Control Assembly. Insert the Display Adjustment PC board (HP part number 85662-60088) into the A3A2 slot. See Figure 3-17 for the location of the A3A2 assembly.

3-46 Adjustments