7.3 MHz Bandwidth Filter Adjustments

25.Adjust A4A7C13 PK for maximum peak-to-peak signal on Channel 2 display. See Figure 3-39 for location of adjustment. If unable to achieve a “peak” in signal amplitude, increase or decrease value of A4A7C12. Refer to Table 3-3 for range of values.

A4A7

3 M H z B a n d w i d t h F i l t e r

Figure 3-39. Location of 3 MHz Peak Adjustments

26.Move Channel 2 probe to A4A7TP3 (left side of C32 SYM).

27.Adjust frequency synthesizer output frequency to peak Channel 1 display.

28.Adjust A4A7C22 PK for maximum peak-to-peak signal on Channel 2 display. See Figure 3-39 for location of adjustment. If unable to achieve a “peak” in signal amplitude, increase or decrease value of A4A7C21. Refer to Table 3-3 for range of values.

29.Move Channel 2 probe to A4A7TPl (left side of C41 SYM).

30.Adjust frequency synthesizer output frequency to peak Channel 1 display.

31.Adjust A4A7C31 PK for maximum peak-to-peak signal on Channel 2 display. See Figure 3-39 for location of adjustment. If unable to achieve a “peak” in signal amplitude, increase or decrease value of A4A7C30. Refer to ‘Iable 3-3 for range of values.

32.Disconnect Channel 2 probe from A4A7TPl.

33.Adjust frequency synthesizer output frequency to peak Channel 1 display.

34.Adjust [REFERENCE LEVEL_) using step keys to place signal near top CRT graticule line.

35.Adjust A4A7C40 PK for maximum signal amplitude on the CRT display. See Figure 3-39 for the location of adjustment. If unable to achieve a “peak” in signal amplitude, increase or decrease value of A4A7C39. Refer to Table 3-3 for range of values.

36.Disconnect Channel 1 probe from A4A7TP7. Disconnect ‘frequency synthesizer output from A4A8Jl and reconnect cable 97 (white/violet).

Adjustments 3-75