4.Final Display Adjustments (SN 3001A and Below)

5.For best overall focusing of the display, adjust the following potentiometers in the sequence listed below:

a.AlA3R14 FOCUS LIMIT for best focus of graticule lines (long vectors)

b.AlA2R36 ASTIG

c.AlA2R30 FOCUS GAIN for best focus of annotation (short vectors)

6.Adjust AlA2R31 ORTHO, the front-panel ALIGN control, and AlA2R32 PATT to optimize the orientation and appearance of the rectangular graticule pattern on the CRT display.

7.Repeat steps 4 through 6 as needed to optimize overall display focus and appearance.

 

 

R27

R31

R32

R36

 

R30

FEN

 

G A I N

O R T H O

P A T T

A S T

I

G

F O C U S G A I N

 

 

\

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AIAZ

Figure 3-30.

Location of Final Display Adjustments on AlA2, AlA4, and AlA

3-60 Adjustments