Tkst 1. Center

Frequency Readout

Accuracy Test

Step 8. Center Frequency Readout Error Test Record

Comb

 

 

 

Spectrum Analyzer

 

 

 

 

Generator

 

 

 

 

 

 

 

 

 

 

 

 

 

Comb

 

 

[FREQUENCY SPAN)

[ C E N T E R FREQUENCY)

T

Center Readout

 

 

Frequency

 

 

 

ww

W-W

 

 

WW

 

 

 

 

 

 

 

Min

Measured

 

 

Max

 

 

 

 

 

 

 

100 MC

 

 

100 MHz

100

98

 

 

 

102

 

 

 

 

 

 

 

 

 

 

 

 

100 MHz

500

498

 

 

 

 

502

 

 

 

100 MHz

1000

998

 

 

 

1002

 

 

 

 

 

EXT TRIG

 

 

10 MHz

100

99.8

 

 

 

100.2

 

 

 

 

 

(1, 2, 5, or 10 MHz)

 

 

10 MHz

500

499.8

 

 

 

500.2

 

 

 

 

 

trigger signal

 

 

10 MHz

1000

999.8

 

 

 

1000.2

 

 

 

 

 

 

 

 

10 MHz

1500

1499.8

 

 

 

1500.2

 

 

 

 

 

 

 

 

 

1 MHz

1000

999.98

 

 

 

1000.02

 

 

 

 

 

 

 

 

 

100 kHz

1000

999.998

 

 

 

1000.002

 

 

 

 

 

 

 

1

 

10 kHz

1000

999.9998

 

 

 

1000.0002

 

 

 

 

 

2-70 Performance Tests

Page 92
Image 92
HP 8568B manual Tkst 1. Center Frequency Readout Accuracy Test