19. Fast Sweep Time Accuracy Test (~20 ms)

19.Fast Sweep Time Accuracy Test (430 ms)

Related Adjustment NoneSpecification &lo% for sweep times 5100 seconds

Description The triangular wave output of a function generator is used to modulate a 500 MHz signal which is applied to the spectrum analyzer SIGNAL INPUT. The signal is demodulated in the zero span mode to display the triangular waveform. Sweep time accuracy for sweep times ~20 ms is tested by checking the spacing of the signal peaks on the displayed waveform.

SPECTRUM ANALYZER

SYNTHESIZED SWEEPER

 

Figure 2-33. Fast Sweep Time Accuracy (~20 ms Test Setup)

Equipment

Function

Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

HP 3312A

 

Universal

Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

HP 5316A

 

Signal Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

HP 8340A

Procedure 1. Connect equipment as shown in Figure 2-33.

2.Press QNSTR PRESET) on spectrum analyzer.3.Key in analyzer settings as follows:

 

 

 

)

500

MHZ

~CENTER F R E Q U E N C Y

 

 

 

1 0 0

kHz

(

F R E Q U E N C Y S P A N

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .)

 

 

 

 

 

 

4.Set synthesized sweeper for an output frequency of 500 MHz and an output power level of -10 dBm.

Performance Tests 2 - 63