8. Frequency Response Test

25.After completion of sweep, press DISPLAY LINE (j?KiK] on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. Record the measurements below.

SIGNAL INPUT 1

(100 kHz to 20 MHz)

MaximumdBm

MinimumdBm

26.Measure and record signal level at start of trace (100 kHz).

SIGNAL INPUT 1

(100 kHz)

dBm

27.Connect output of frequency synthesizer to SIGNAL INPUT 2. Activate this input with the pushbutton.

Option 001. Do not use HP 11852A Minimum Loss Pad. Set frequency synthesizer output amplitude to -2 dBm.

28.Press TRACE A CCLEAR-WRITE)and DISPLAY LINE (OFF) on spectrum analyzer.

29.Set frequency synthesizer FREQUENCY to 20 MHz. Set spectrum analyzer (CENTER FREQUENCY_) to 20 MHz, and (FREQUENCY SPAN) to

1MHz.

30.Repeat steps 22 through 24.

31.After completion of sweep, press DISPLAY LINE [ENTER] on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. Record the measurements below.

SIGNAL INPUT 2

(100 kHz to 20 MHz)

MaximumdBm

MinimumdBm

Figure 2-15.

Frequency Response Test Setup (100 Hz to 100 k=)

P e r f o r m a n c e T e s t s 2-27