6.Impulse and Resolution Bandwidth Switching Uncertainty Tkst

Related Adjustment 3 MHz Bandwidth Filter Adjustments

21.4Bandwidth Filter Adjustments Down/Up Converter Adjustments

Specification f2.0 dB, 10 Hz bandwidth f0.8 dB, 30 Hz bandwidth

f0.5 dB, 100 Hz to 1 MHz bandwidth * 1 .O dB, 3 MHz bandwidth

30 kHz and 100 kHz bandwidth switching uncertainty figures only applicable 190% Relative Humidity.

Description The CAL OUTPUT signal is applied to the input of the spectrum analyzer. The deviation in peak amplitude of the signal trace is then measured as each resolution bandwidth filter is switched in.

Equipment None required

Procedure 1. Press ~NSTR PRESET).

2.Connect CAL OUTPUT to SIGNAL INPUT 2.

3.Key in the following control settings:

(

CENTER FREQUENCY]

 

 

 

.20

MHz

 

 

 

 

 

.

.

5

MHz

FREQUENCY S P A N

) .

 

 

 

 

 

 

 

 

 

. - 8

dBm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

. . . . . .

1

MHz

4. Press LOG [ENTER dB/otv_) and key in 1 dB. Press MARKER

(PEAK SEARCH) a].

5.Key in settings according to Table 4-6. Press MARKER

[PEAK SEARCH) at each setting, then read the amplitude deviation from the MARKER [nl readout at the upper right of the display (see Figure 4-6).The allowable deviation for each resolution bandwidth setting is shown in the table.

4-16 Option 462

Page 281
Image 281
HP 8568B manual Procedure 1. Press ~NSTR Preset, Press LOG Enter dB/otv and key in 1 dB. Press Marker