4. Final Display Adjustments (SN 3004A and Above)

4.Final Display Adjustments (SN 3004A and Above)

Reference AlAl KeyboardAlA X, Y, Z Axis Amplifiers

Description This procedure is used to optimize the appearance of the CRT display during routine maintenance or after CRT replacement or minor repairs. First, the display is adjusted for best focus over the full CRT, then the graticule pattern is adjusted for optimum rectangular display.

Equipment Digital Photometer . . . . . . . . . . . . . . . . . . . . . . . . .Tektronix J-16, Option 02 Photometer Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Tektronix 56503 Photometer interconnect cable . . . . . . . . . . . . . . . Tektronix 012-0414-02 Photometer light occluder . . . . . . . . . . . . . . . . . . . . .Tektronix 016-0305-00

Procedure Note

Adjustment Procedure 2, High Voltage Adjustment (SN 3004A and Above) should be performed prior to performing the following adjustment procedure.

1. Connect the equipment as shown in Figure 3-31.

SPECTRUl ANALYZER

Figure 3-31. Final Display Adjustments Setup

2.Set the photometer probe to NORMAL. Press [POWER_) on the photometer to turn it on and allow 30 minutes warm-up. Zero the photometer according to the manufacturer’s instructions.

3.With the spectrum analyzer’s LINE switch set to STANDBY, set the potentiometers listed in the ‘Iable 3-6 as indicated. See Figure 3-32 for the location of the adjustments.

Note In this procedure, do not adjust the following potentiometers and variable capacitors on the AlA X, Y, Z Amplifier Assembly: C104, C109, C204, C209, C307, R117, R217, or R308. These components are adjusted in the factory and in Adjustment Procedure 3, Preliminary Display Adjustments (SN 3004A and Above).

Adjustments 3.61